Assessment of Radiation Dose Received by Orthopedic and Spine Operating Room Personnel during C-Arm Use Using Experimental Methods and Monte Carlo Simulation

Introduction: Due to increasing use of C-arm in surgeries, personnel are exposed to a greater risk of radiation. Accurate assessment of the dose received during various surgeries is one of the most important measures to prevent radiation-related complications. This study aimed to evaluate the dose received by operating room staff in orthopedic and spinal surgeries while using the C-arm device utilizing experimental methods and Monte Carlo simulation. Method: This cross-sectional study was conducted from September to February 2023. To record the absorbed dose, TLD 100 dosimeters were used, and instead of humans, phantoms were utilized. The dosimeter was placed on the thyroid of the phantom. The phantoms were positioned both behind and in front of a lead wall at three locations: right, center, and left. The C-arm device was used to simulate orthopedic and neurosurgery procedures in both anteroposterior and lateral positions. Data analysis was performed using SPSS software (version 24), with a significance level set at 0.05 for the tests. Results: The absorbed dose for the phantoms positioned behind the lead wall was significantly lower than that in front of the wall (P<0.05). When the phantom was in front of the wall, it received a higher dose in the lateral position of the C-arm, compared to the anteroposterior position (P<0.05). Additionally, when the phantom was placed behind the lead wall in the center and left positions, it received a higher dose in the anteroposterior position of the C-arm, compared to the lateral position (P<0.05). Conclusion: The results of this study indicated a significant impact of the C-arm device's position, as well as the use of protective equipment on the dose received by operating room personnel during orthopedic and neurosurgical procedures. It emphasizes the importance of using appropriate shields when working with the C-arm.

Institutions

Publication Details

Journal
Journal of North Khorasan University of Medical Sciences
Published
2024-12-31
DOI
https://doi.org/10.32592/nkums.17.2.75
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Assessment of Radiation Dose Received by Orthopedic and Spine Operating Room Personnel during C-Arm Use Using Experimental Methods and Monte Carlo Simulation

Journal of North Khorasan University of Medical Sciences
Radiation Dose and Imaging
article

Assessment of Radiation Dose Received by Orthopedic and Spine Operating Room Personnel during C-Arm Use Using Experimental Methods and Monte Carlo Simulation

article en

Abstract

Introduction: Due to increasing use of C-arm in surgeries, personnel are exposed to a greater risk of radiation. Accurate assessment of the dose received during various surgeries is one of the most important measures to prevent radiation-related complications. This study aimed to evaluate the dose received by operating room staff in orthopedic and spinal surgeries while using the C-arm device utilizing experimental methods and Monte Carlo simulation. Method: This cross-sectional study was conducted from September to February 2023. To record the absorbed dose, TLD 100 dosimeters were used, and instead of humans, phantoms were utilized. The dosimeter was placed on the thyroid of the phantom. The phantoms were positioned both behind and in front of a lead wall at three locations: right, center, and left. The C-arm device was used to simulate orthopedic and neurosurgery procedures in both anteroposterior and lateral positions. Data analysis was performed using SPSS software (version 24), with a significance level set at 0.05 for the tests. Results: The absorbed dose for the phantoms positioned behind the lead wall was significantly lower than that in front of the wall (P<0.05). When the phantom was in front of the wall, it received a higher dose in the lateral position of the C-arm, compared to the anteroposterior position (P<0.05). Additionally, when the phantom was placed behind the lead wall in the center and left positions, it received a higher dose in the anteroposterior position of the C-arm, compared to the lateral position (P<0.05). Conclusion: The results of this study indicated a significant impact of the C-arm device's position, as well as the use of protective equipment on the dose received by operating room personnel during orthopedic and neurosurgical procedures. It emphasizes the importance of using appropriate shields when working with the C-arm.

Journal of North Khorasan University of Medical Sciences
Hamedan University of Medical Sciences (IR)
Openalex Percentile: Top 55%
Radiation Dose and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.