The validity of the VO<sub>2</sub> Master Pro for measuring oxygen consumption during sedentary activity and treadmill walking and jogging

Metabolic carts are commonly used to estimate oxygen consumption (V̇O<sub>2</sub>) during exercise, but are largely limited to controlled laboratory settings. The V̇O<sub>2</sub> Master Pro (VMP) is a newer, portable metabolic analyzer designed to address this limitation; however, few studies have evaluated the validity of this device at varying activity levels. This study aimed to assess the validity of the VMP in measuring V̇O<sub>2</sub> compared with a stationary metabolic cart, the COSMED Quark CPET (CQ), during sedentary activity and treadmill walking/jogging in a laboratory setting. Twenty-seven healthy adults (mean age = 22.1 ± 7.6 years; female = 51.8%) participated in two laboratory trials on separate days. In a counterbalanced order, participants used the CQ and VMP during 10 min conditions of the following activities: sedentary activity (sitting quietly), slow walking (3.2 km/h), brisk walking (5.6 km/h), and jogging (7.2 km/h). The agreement between the two measures was evaluated using equivalence testing, mean absolute percentage error (MAPE), percentage bias, intraclass correlation coefficients (ICCs), and Bland-Altman analyses. The devices showed low agreement and significant proportional biases across all activity levels (ICCs = 0.135-0.323). Equivalence testing did not demonstrate statistically significant equivalence between the devices (<i>p</i> > 0.05), with the VMP underestimating V̇O<sub>2</sub>. The smallest error appeared during jogging (MAPE = 20.05%; percentage bias= -19.29%). The VMP underestimated V̇O<sub>2</sub> at all tested intensities, demonstrating low accuracy and agreement relative to the reference measure. This may be attributed to limited ventilatory flow capture or sensor responsiveness during submaximal activities. Observed bias and within-subject variability suggest caution when using the VMP across different submaximal activity levels.

Authors

Institutions

Publication Details

Journal
Applied Physiology Nutrition and Metabolism
Published
2025-01-01
DOI
https://doi.org/10.1139/apnm-2025-0095
Primary Topic
Cardiovascular and exercise physiology
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The validity of the VO2 Master Pro for measuring oxygen consumption during sedentary activity and treadmill walking and jogging

Xiaolin Wen, Youngdeok Kim, Kyu-Ri Hong, Junhao Shen et al.
1 links
Applied Physiology Nutrition and Metabolism
Cardiovascular and exercise physiology
article

The validity of the VO2 Master Pro for measuring oxygen consumption during sedentary activity and treadmill walking and jogging

Xiaolin Wen, Youngdeok Kim, Kyu-Ri Hong, Junhao Shen, Ronald K. Evans, Jonathan Kenyon, In-Whi Hwang, Ho-Jun Kim, Jessica Geller, Jisu Kim, Jung Min Lee
article en

Abstract

Metabolic carts are commonly used to estimate oxygen consumption (V̇O<sub>2</sub>) during exercise, but are largely limited to controlled laboratory settings. The V̇O<sub>2</sub> Master Pro (VMP) is a newer, portable metabolic analyzer designed to address this limitation; however, few studies have evaluated the validity of this device at varying activity levels. This study aimed to assess the validity of the VMP in measuring V̇O<sub>2</sub> compared with a stationary metabolic cart, the COSMED Quark CPET (CQ), during sedentary activity and treadmill walking/jogging in a laboratory setting. Twenty-seven healthy adults (mean age = 22.1 ± 7.6 years; female = 51.8%) participated in two laboratory trials on separate days. In a counterbalanced order, participants used the CQ and VMP during 10 min conditions of the following activities: sedentary activity (sitting quietly), slow walking (3.2 km/h), brisk walking (5.6 km/h), and jogging (7.2 km/h). The agreement between the two measures was evaluated using equivalence testing, mean absolute percentage error (MAPE), percentage bias, intraclass correlation coefficients (ICCs), and Bland-Altman analyses. The devices showed low agreement and significant proportional biases across all activity levels (ICCs = 0.135-0.323). Equivalence testing did not demonstrate statistically significant equivalence between the devices (<i>p</i> > 0.05), with the VMP underestimating V̇O<sub>2</sub>. The smallest error appeared during jogging (MAPE = 20.05%; percentage bias= -19.29%). The VMP underestimated V̇O<sub>2</sub> at all tested intensities, demonstrating low accuracy and agreement relative to the reference measure. This may be attributed to limited ventilatory flow capture or sensor responsiveness during submaximal activities. Observed bias and within-subject variability suggest caution when using the VMP across different submaximal activity levels.

Applied Physiology Nutrition and MetabolismVol. 50
Virginia Commonwealth University (US), Kyung Hee University (KR)
Clean water and sanitation
Openalex Percentile: Top 73%
Cardiovascular and exercise physiology
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.