A discussion on two alternative approaches to CT coupon based K1C determination providing results comparable with ASTM E 399

The value of Plane Strain Fracture Toughness of a material calculated in accordance with the ASTM E 399 has been noted to be a specimen size-dependent value. The size-dependency of the same with respect to ductile materials has been an area of debate both for quality testing in production and the best applicable value for new structural design of components. While in case of production quality assessment the primary challenge is the repeatability of complying results for ductile materials, for new design – it is of a representative value for the material. Crack Mouth Opening Displacement (CMOD) and the applied load are traditionally used to calculate K 1C from CT specimen by ASTM (2023). A discussion on the comparison of two K 1C equivalents is presented here based upon observations with respect to specific Aluminium alloys and their relationship with this established material indicator is explored. Equivalents considered are K 1si and round tensile specimen inspired by Sarchamy et al (1996) and Wilson (1997). While K 1si is based upon CMOD estimation by assuming a fixed amount of crack extension of 0.5 mm irrespective of tested specimen size, round specimen may be applicable when dimensionally constrained.

Authors

Institutions

Publication Details

Journal
Procedia Structural Integrity
Published
2025-01-01
DOI
https://doi.org/10.1016/j.prostr.2025.06.080
Primary Topic
Nuclear Physics and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A discussion on two alternative approaches to CT coupon based K1C determination providing results comparable with ASTM E 399

Mr. Parthasarathy Iyengar, Mr. Jon Mardaras, Shwe Soe
Procedia Structural Integrity
Nuclear Physics and Applications
article

A discussion on two alternative approaches to CT coupon based K1C determination providing results comparable with ASTM E 399

Mr. Parthasarathy Iyengar, Mr. Jon Mardaras, Shwe Soe
article en

Abstract

The value of Plane Strain Fracture Toughness of a material calculated in accordance with the ASTM E 399 has been noted to be a specimen size-dependent value. The size-dependency of the same with respect to ductile materials has been an area of debate both for quality testing in production and the best applicable value for new structural design of components. While in case of production quality assessment the primary challenge is the repeatability of complying results for ductile materials, for new design – it is of a representative value for the material. Crack Mouth Opening Displacement (CMOD) and the applied load are traditionally used to calculate K 1C from CT specimen by ASTM (2023). A discussion on the comparison of two K 1C equivalents is presented here based upon observations with respect to specific Aluminium alloys and their relationship with this established material indicator is explored. Equivalents considered are K 1si and round tensile specimen inspired by Sarchamy et al (1996) and Wilson (1997). While K 1si is based upon CMOD estimation by assuming a fixed amount of crack extension of 0.5 mm irrespective of tested specimen size, round specimen may be applicable when dimensionally constrained.

Procedia Structural IntegrityVol. 68
Airbus (France) (FR), University of the West of England (GB), Airbus (United Kingdom) (GB)
Openalex Percentile: Top 76%
Nuclear Physics and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.