cBP-Tnet: Continuous Blood Pressure Estimation Using Multi-Task Transformer Network with Automatic Photoplethysmogram Feature Extraction

Traditional cuff-based blood pressure (BP) monitoring methods provide only intermittent readings, while invasive alternatives pose clinical risks. Recent studies have demonstrated feasibility of estimating continuous non-invasive cuff-less BP using photoplethysmogram (PPG) signals alone. However, existing approaches rely on complex manual feature engineering and/or multiple model architectures, resulting in inefficient epoch training numbers and limited performance. This research proposes cBP-Tnet, an efficient single-channel and model multi-task Transformer network designed for PPG signal automatic feature extraction. cBP-Tnet employed specialized hyperparameters—integrating adaptive Kalman filtering, outlier elimination, signal synchronization, and data augmentation—leveraging multi-head self-attention and multi-task learning strategies to identify subtle and shared waveform patterns associated with systolic blood pressure (SBP) and diastolic blood pressure (DBP). We used the MIMIC-II public dataset (500 patients with 202,956 samples) for experimentation. Results showed mean absolute errors of 4.32 mmHg for SBP and 2.18 mmHg for DBP. For the first time, both SBP and DBP meet the Association for the Advancement of Medical Instrumentation’s international standard (<5 mmHg, >85 subjects). Furthermore, the network efficiently reduces the epoch training number by 13.67% when compared to other deep learning methods. Thus, this establishes cBP-Tnet’s potential for integration into wearable and home-based healthcare devices with continuous non-invasive cuff-less blood pressure monitoring.

Authors

Institutions

Publication Details

Journal
Applied Sciences
Published
2025-07-12
DOI
https://doi.org/10.3390/app15147824
Citations
2
Primary Topic
Non-Invasive Vital Sign Monitoring
Type
article
Field-Weighted Citation Impact
1.35

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

cBP-Tnet: Continuous Blood Pressure Estimation Using Multi-Task Transformer Network with Automatic Photoplethysmogram Feature Extraction

Angelino A. Pimentel, Ji-Jer Huang, Aaron Raymond See
2 citations
Applied Sciences
Non-Invasive Vital Sign Monitoring
1.35
article

cBP-Tnet: Continuous Blood Pressure Estimation Using Multi-Task Transformer Network with Automatic Photoplethysmogram Feature Extraction

Angelino A. Pimentel, Ji-Jer Huang, Aaron Raymond See
article en
2 citations

Abstract

Traditional cuff-based blood pressure (BP) monitoring methods provide only intermittent readings, while invasive alternatives pose clinical risks. Recent studies have demonstrated feasibility of estimating continuous non-invasive cuff-less BP using photoplethysmogram (PPG) signals alone. However, existing approaches rely on complex manual feature engineering and/or multiple model architectures, resulting in inefficient epoch training numbers and limited performance. This research proposes cBP-Tnet, an efficient single-channel and model multi-task Transformer network designed for PPG signal automatic feature extraction. cBP-Tnet employed specialized hyperparameters—integrating adaptive Kalman filtering, outlier elimination, signal synchronization, and data augmentation—leveraging multi-head self-attention and multi-task learning strategies to identify subtle and shared waveform patterns associated with systolic blood pressure (SBP) and diastolic blood pressure (DBP). We used the MIMIC-II public dataset (500 patients with 202,956 samples) for experimentation. Results showed mean absolute errors of 4.32 mmHg for SBP and 2.18 mmHg for DBP. For the first time, both SBP and DBP meet the Association for the Advancement of Medical Instrumentation’s international standard (<5 mmHg, >85 subjects). Furthermore, the network efficiently reduces the epoch training number by 13.67% when compared to other deep learning methods. Thus, this establishes cBP-Tnet’s potential for integration into wearable and home-based healthcare devices with continuous non-invasive cuff-less blood pressure monitoring.

Applied SciencesVol. 15(14)
Saint Mary's University (PH), National Chin-Yi University of Technology (TW), Nueva Vizcaya State University (PH), Southern Taiwan University of Science and Technology (TW)
Ministry of Science and Technology, Taiwan
Openalex Percentile: Top 19%
Non-Invasive Vital Sign Monitoring
1.35
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.