Analysis of X-ray Fluoroscopy (Siemens) Quality control Using X-ray Multimeter (R/F Raysafe) and PMMA Phantom

Abstract. Diagnostic and interventional radiology X-ray machines used by health facilities are required to undergo periodic testing before being put into service. This study was conducted to analyze the appropriateness of a fluoroscopic X-ray machine at a type A hospital in Banda Aceh. Direct quality assurance measurements were performed using a PMMA-type phantom, a fluoroscopic X-ray machine (SIEMENS AGILE MAX), X-ray multimeters, and aluminum filters. The BAPETEN standard No. 2 in 2018 was used as a reference. Based on the measurement analysis results, the voltage accuracy measurement is deviated by 0.13–0.3% higher from the voltage setting; however, it remained within the minimum threshold by BAPETEN standard. The radiation beam quality value (HVL) is relatively low compared to the standard minimum value and remained below the threshold. The typical dose rate has met the standard, and the maximum dose rate in the air is also below the set standard. However, the maximum dose rate in the air was observed to be 60 mGy/min at a voltage of 100 kVp.

Authors

Publication Details

Journal
Journal of Aceh Physics Society
Published
2025-04-14
DOI
https://doi.org/10.24815/jacps.v14i1.44604
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Analysis of X-ray Fluoroscopy (Siemens) Quality control Using X-ray Multimeter (R/F Raysafe) and PMMA Phantom

Rini Safitri, Elin Yusibani, Teuku Khairuman, Fashbir Fashbir et al.
Journal of Aceh Physics Society
Advanced X-ray and CT Imaging
article

Analysis of X-ray Fluoroscopy (Siemens) Quality control Using X-ray Multimeter (R/F Raysafe) and PMMA Phantom

Rini Safitri, Elin Yusibani, Teuku Khairuman, Fashbir Fashbir, Umiatin Umiatin, Kurnia Lahna, Husnul Hayati
article en

Abstract

Abstract. Diagnostic and interventional radiology X-ray machines used by health facilities are required to undergo periodic testing before being put into service. This study was conducted to analyze the appropriateness of a fluoroscopic X-ray machine at a type A hospital in Banda Aceh. Direct quality assurance measurements were performed using a PMMA-type phantom, a fluoroscopic X-ray machine (SIEMENS AGILE MAX), X-ray multimeters, and aluminum filters. The BAPETEN standard No. 2 in 2018 was used as a reference. Based on the measurement analysis results, the voltage accuracy measurement is deviated by 0.13–0.3% higher from the voltage setting; however, it remained within the minimum threshold by BAPETEN standard. The radiation beam quality value (HVL) is relatively low compared to the standard minimum value and remained below the threshold. The typical dose rate has met the standard, and the maximum dose rate in the air is also below the set standard. However, the maximum dose rate in the air was observed to be 60 mGy/min at a voltage of 100 kVp.

Journal of Aceh Physics SocietyVol. 14(1)
Openalex Percentile: Top 84%
Advanced X-ray and CT Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.