Image Quality Assessment of In-House Phantom for Pediatric Computed Tomography

The study aimed to determine the estimated low dose and image quality of the in-house pediatric phantoms using the 128 slice CT-Scan. This study used in house pediatric phantoms (7 year). Twelve artificial cylinder targets were fabricated with diameters 3, 5, 8, and 10 mm, and CT densities -+100, -400, -750 HU. The image quality was performed using CT-Scan 128 slice (GE) with parameters of helical scanning using a low dose and standard dose and using 80 and 120 kV. For Signal noise to ratio (SNR), Contrast to Noise Ratio (CNR) were also investigated. Volumetric CT dose index (CTDIvol), Size‐Specific Dose Estimates (SSDE). The effective value of the diameter for 7-year-old was 20.15 cm. According to the study, the amount of radiation exposure measured as SSDE for in-house pediatric phantoms was found to be 3.71 mGy for low dose and 6.09 for standard dose. For image quality analysis, the low-dose CT protocol has a detection sensitivity of 100% for targets measuring > 5mm in diameter.

Authors

Institutions

Publication Details

Journal
Jurnal Fisika dan Aplikasinya
Published
2025-06-30
DOI
https://doi.org/10.12962/j24604682.v21i2.20434
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Image Quality Assessment of In-House Phantom for Pediatric Computed Tomography

Endarko Endarko, Aditya Prayugo Hariyanto, Dafa Miftahuddin, Audiena Gelung Prayitno et al.
Jurnal Fisika dan Aplikasinya
Radiation Dose and Imaging
article

Image Quality Assessment of In-House Phantom for Pediatric Computed Tomography

Endarko Endarko, Aditya Prayugo Hariyanto, Dafa Miftahuddin, Audiena Gelung Prayitno, M R A Gani
article en

Abstract

The study aimed to determine the estimated low dose and image quality of the in-house pediatric phantoms using the 128 slice CT-Scan. This study used in house pediatric phantoms (7 year). Twelve artificial cylinder targets were fabricated with diameters 3, 5, 8, and 10 mm, and CT densities -+100, -400, -750 HU. The image quality was performed using CT-Scan 128 slice (GE) with parameters of helical scanning using a low dose and standard dose and using 80 and 120 kV. For Signal noise to ratio (SNR), Contrast to Noise Ratio (CNR) were also investigated. Volumetric CT dose index (CTDIvol), Size‐Specific Dose Estimates (SSDE). The effective value of the diameter for 7-year-old was 20.15 cm. According to the study, the amount of radiation exposure measured as SSDE for in-house pediatric phantoms was found to be 3.71 mGy for low dose and 6.09 for standard dose. For image quality analysis, the low-dose CT protocol has a detection sensitivity of 100% for targets measuring > 5mm in diameter.

Jurnal Fisika dan AplikasinyaVol. 21(2)
Sepuluh Nopember Institute of Technology (ID), Dharmais Cancer Hospital (ID)
Institut Teknologi Sepuluh Nopember
Openalex Percentile: Top 75%
Radiation Dose and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.