Noninvasive Analysis of Biological Components Using Simplified Mid-Infrared Photothermal Deflection Spectroscopy

We developed a photothermal deflection spectroscopy (PTDS) system for the noninvasive analysis of biological tissue. This system detects heat induced by irradiation with pulse-modulated mid-infrared light as the deflection of a probe laser. The probe light is incident on the sensing element horizontal with respect to its contact surface with the sample. This setup simplifies the optical alignment compared to conventional systems, which require the probe laser to be totally reflected at the prism contact surface and aligned with the point of mid-infrared light irradiation. In this study, we measured the PTDS spectra of biological samples to determine the characteristic features of their infrared absorption. We also compared the measurement reproducibility of two configurations: a horizontal optical path and a total reflection optical path. The horizontal optical path showed greater measurement reproducibility than the total reflection optical path when performing intermittent measurements on the wrist.

Authors

Institutions

Publication Details

Journal
Sensors
Published
2025-07-12
DOI
https://doi.org/10.3390/s25144368
Citations
1
Primary Topic
Thermography and Photoacoustic Techniques
Type
article
Field-Weighted Citation Impact
0.86

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Noninvasive Analysis of Biological Components Using Simplified Mid-Infrared Photothermal Deflection Spectroscopy

Saiko Kino, Yuji Matsuura, Hiroto Ito
1 citations
Sensors
Thermography and Photoacoustic Techniques
0.86
article

Noninvasive Analysis of Biological Components Using Simplified Mid-Infrared Photothermal Deflection Spectroscopy

Saiko Kino, Yuji Matsuura, Hiroto Ito
article en
1 citations

Abstract

We developed a photothermal deflection spectroscopy (PTDS) system for the noninvasive analysis of biological tissue. This system detects heat induced by irradiation with pulse-modulated mid-infrared light as the deflection of a probe laser. The probe light is incident on the sensing element horizontal with respect to its contact surface with the sample. This setup simplifies the optical alignment compared to conventional systems, which require the probe laser to be totally reflected at the prism contact surface and aligned with the point of mid-infrared light irradiation. In this study, we measured the PTDS spectra of biological samples to determine the characteristic features of their infrared absorption. We also compared the measurement reproducibility of two configurations: a horizontal optical path and a total reflection optical path. The horizontal optical path showed greater measurement reproducibility than the total reflection optical path when performing intermittent measurements on the wrist.

SensorsVol. 25(14)
Tohoku University (JP)
Japan Society for the Promotion of Science
Openalex Percentile: Top 28%
Thermography and Photoacoustic Techniques
0.86
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.