Pulsed electric field ablation process-the effect of bifurcation stents on electric field and heat distribution

Pulsed field ablation (PFA) is a novel non-thermal technique for arrhythmia treatment, capable of targeting the epicardium. However, the effects of coronary bifurcation lesions (BL) and bifurcation stents (BS) on ablation efficacy have not been evaluated. A three-dimensional computer simulation model was developed to assess the effect of the ablation region. The ablation catheter was set to be located directly above the BS model by performing a true size reduction of the ablation device. The ablation region was evaluated using an electric field strength contour of 1000 V/cm. The influence of coronary BL and BS on PFA ablation efficacy was assessed using ablation effect maps from different models. BS did not induce electric field coupling, and the distortion of the electric field in the ablation region remained unchanged, with cold and hot spots located similarly. The presence of coronary branches reduced the ablation area width, with a maximum difference of 2.82 mm. The angle between the ablation catheter and the main coronary artery branch had a smaller effect on the ablation zone width, with a maximum difference of 1.24 mm. The presence of BS will have a positive effect on the temperature increase in the ablation region (3.59 °C temperature increase in the presence of BS). The results indicate that BL and BS influence the electric field distribution in the ablation region. However, this distortion is confined to the ablation area, with no coupling between the main coronary artery and its branches to exacerbate the electric field distortion. The presence of BS exacerbates the temperature rise at the ablation site, but this increase does not cause thermal damage to the region.

Authors

Institutions

Publication Details

Journal
Scientific Reports
Published
2025-07-14
DOI
https://doi.org/10.1038/s41598-025-10606-w
Primary Topic
Electrohydrodynamics and Fluid Dynamics
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Pulsed electric field ablation process-the effect of bifurcation stents on electric field and heat distribution

Ming Liang, Yong Yan, Lisheng Xu, Yaling Han et al.
Scientific Reports
Electrohydrodynamics and Fluid Dynamics
article

Pulsed electric field ablation process-the effect of bifurcation stents on electric field and heat distribution

Ming Liang, Yong Yan, Lisheng Xu, Yaling Han, Zhen Wang, Jie Zhang, Jingyang Sun, Fengqi Xuan
article en

Abstract

Pulsed field ablation (PFA) is a novel non-thermal technique for arrhythmia treatment, capable of targeting the epicardium. However, the effects of coronary bifurcation lesions (BL) and bifurcation stents (BS) on ablation efficacy have not been evaluated. A three-dimensional computer simulation model was developed to assess the effect of the ablation region. The ablation catheter was set to be located directly above the BS model by performing a true size reduction of the ablation device. The ablation region was evaluated using an electric field strength contour of 1000 V/cm. The influence of coronary BL and BS on PFA ablation efficacy was assessed using ablation effect maps from different models. BS did not induce electric field coupling, and the distortion of the electric field in the ablation region remained unchanged, with cold and hot spots located similarly. The presence of coronary branches reduced the ablation area width, with a maximum difference of 2.82 mm. The angle between the ablation catheter and the main coronary artery branch had a smaller effect on the ablation zone width, with a maximum difference of 1.24 mm. The presence of BS will have a positive effect on the temperature increase in the ablation region (3.59 °C temperature increase in the presence of BS). The results indicate that BL and BS influence the electric field distribution in the ablation region. However, this distortion is confined to the ablation area, with no coupling between the main coronary artery and its branches to exacerbate the electric field distortion. The presence of BS exacerbates the temperature rise at the ablation site, but this increase does not cause thermal damage to the region.

Scientific ReportsVol. 15(1)
General Hospital of Shenyang Military Region (CN), Tianjin Chest Hospital (CN), Northeastern University (CN)
Affordable and clean energy
Openalex Percentile: Top 85%
Electrohydrodynamics and Fluid Dynamics
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.