Different types of Gafchromic films, for radiotherapy use, are recommended for different dose ranges. Ashland Specialty Ingredients has aimed to continuously develop its products to improve their practical application. Thus, EBT3 was replaced by EBT4, intended to provide better signal to noise ratio (SNR); while MD-V3 was introduced for use at higher dose ranges, in addition to EBT-XD. At present there are limited studies on MD-V3. This study aimed to investigate some relevant characteristics of EBT4 and MD-V3, compared with those of EBT3 and EBT-XD. The parameters investigated were dose response, optical density change with post-irradiation time, orientation effect, signal to noise ratio, polarisation, and lateral response artefact (LRA). EBT4 is similar to EBT3 however it provides better SNR and larger response change with post-irradiation time. EBT-XD and MD-V3 are recommended by the suppliers for high dose range, although the sensitivity curves show that EBT3 and EBT4 could also be used for relatively high dose ranges. All films have orientation effects, with EBT3 the worst. An important characteristic of MD-V3 is that the LRA remains similar, irrespective of delivered dose. These comparative characteristics are intended to be informative for clinical practice involving Gafchromic film use in high dose therapy applications. Recommendations from this study are to use EBT4 for dosimetry in lower-dose applications, provided that both calibration and clinical timings post-irradiation are kept similar, while MD-V3 is the preferred film for high-dose procedures.

Authors

Institutions

Publication Details

Journal
Physical and Engineering Sciences in Medicine
Published
2025-07-14
DOI
https://doi.org/10.1007/s13246-025-01596-0
Citations
1
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
1.88

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparative characterisation of different types of Gafchromic films for radiotherapy use

Martin J Butson, Nick Bennie, David Thwaites, Tarafder Shameem
1 citations
Physical and Engineering Sciences in Medicine
Advanced Radiotherapy Techniques
1.88
article

Comparative characterisation of different types of Gafchromic films for radiotherapy use

Martin J Butson, Nick Bennie, David Thwaites, Tarafder Shameem
article en
1 citations

Abstract

Different types of Gafchromic films, for radiotherapy use, are recommended for different dose ranges. Ashland Specialty Ingredients has aimed to continuously develop its products to improve their practical application. Thus, EBT3 was replaced by EBT4, intended to provide better signal to noise ratio (SNR); while MD-V3 was introduced for use at higher dose ranges, in addition to EBT-XD. At present there are limited studies on MD-V3. This study aimed to investigate some relevant characteristics of EBT4 and MD-V3, compared with those of EBT3 and EBT-XD. The parameters investigated were dose response, optical density change with post-irradiation time, orientation effect, signal to noise ratio, polarisation, and lateral response artefact (LRA). EBT4 is similar to EBT3 however it provides better SNR and larger response change with post-irradiation time. EBT-XD and MD-V3 are recommended by the suppliers for high dose range, although the sensitivity curves show that EBT3 and EBT4 could also be used for relatively high dose ranges. All films have orientation effects, with EBT3 the worst. An important characteristic of MD-V3 is that the LRA remains similar, irrespective of delivered dose. These comparative characteristics are intended to be informative for clinical practice involving Gafchromic film use in high dose therapy applications. Recommendations from this study are to use EBT4 for dosimetry in lower-dose applications, provided that both calibration and clinical timings post-irradiation are kept similar, while MD-V3 is the preferred film for high-dose procedures.

Physical and Engineering Sciences in MedicineVol. 48(3)
The University of Sydney (AU), New South Wales Department of Health (AU), Taronga Conservation Society Australia (AU), Northern Sydney Local Health District (AU)
University of Sydney
Openalex Percentile: Top 15%
Advanced Radiotherapy Techniques
1.88
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.