Objective: to conduct bench tests and determine the working range of the pump speed for an implanted left ven- tricular bypass system aimed at diagnosing and treating patients with low anthropometric status. Materials and methods. The axial pump was investigated using a custom-developed hydrodynamic test bench simulating the cardiovascular system. The bench included systems for pressure and flow measurement and registration, along with software for processing both technical and biomedical parameters. Results. The operating range of the rotor speed for the STREAM CARDIO pump required to achieve a flow rate of 2.5 ± 0.5 L/min at a pressure drop of 80 ± 5 mm Hg is 8000 ± 1000 rpm, with a power consumption of 6.5 ± 1 W.

Authors

Institutions

Publication Details

Journal
Russian Journal of Transplantology and Artificial Organs
Published
2025-07-12
DOI
https://doi.org/10.15825/1995-1191-2025-2-81-88
Primary Topic
Mechanical Circulatory Support Devices
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Bench studies of a small axial pump for implantation in patients with low anthropometry

G. А. Shevchenko, N. V. Grudinin, А. П. Кулешов, А. А. Дробышев et al.
Russian Journal of Transplantology and Artificial Organs
Mechanical Circulatory Support Devices
article

Bench studies of a small axial pump for implantation in patients with low anthropometry

G. А. Shevchenko, N. V. Grudinin, А. П. Кулешов, А. А. Дробышев, А. С. Бучнев, V. A. Elenkin
article en

Abstract

Objective: to conduct bench tests and determine the working range of the pump speed for an implanted left ven- tricular bypass system aimed at diagnosing and treating patients with low anthropometric status. Materials and methods. The axial pump was investigated using a custom-developed hydrodynamic test bench simulating the cardiovascular system. The bench included systems for pressure and flow measurement and registration, along with software for processing both technical and biomedical parameters. Results. The operating range of the rotor speed for the STREAM CARDIO pump required to achieve a flow rate of 2.5 ± 0.5 L/min at a pressure drop of 80 ± 5 mm Hg is 8000 ± 1000 rpm, with a power consumption of 6.5 ± 1 W.

Russian Journal of Transplantology and Artificial OrgansVol. 27(2)
Peoples' Friendship University of Russia (RU), V.I.Shumakov Federal Research Center of Transplantology and Artificial Organs (RU)
Openalex Percentile: Top 84%
Mechanical Circulatory Support Devices
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Bench studies of a small axial pump for implantation in patients with low anthropometry — G. А. Shevchenko, N. V. Grudinin, et al. · Russian Journal of Transplantology and Artificial Organs (2025) | TGRS Research Map | TGRS