Enhancing Imaging Performance and Resolution in Magneto‐Acoustic Electrical Tomography With Magnetic Field Measurements ( <scp>MAET</scp> ‐ <scp>MI</scp> ) Using Figure‐of‐Eight and High‐Quality Factor Circular Coils

Magneto-acousto-electrical tomography with magnetic field measurement technique (MAET-MI) is a hybrid imaging method that brings high spatial resolution of ultrasound imaging in electrical impedance tomography. This study investigates the impact of the quality factor of circular and figure-of-eight coils on the imaging performance of MAET-MI. Induced MAET signals on the circular coil are accurately obtained by modeling a circuit representation of an air-cored circular coil and deriving its transfer function through impedance measurements. The study demonstrates a significant improvement in signal-to-noise ratio (SNR) using high-quality factor coils compared to unity quality factor coils. Additionally, a 16-element linear phased array (LPA) ultrasound transducer, an air core circular coil, and a figure-of-eight coil are numerically modeled to obtain sector scan images of two-dimensional conductivity distributions. Point spread function (PSF) is characterized, and the lateral resolution of sector scan conductivity images is enhanced through two-dimensional deconvolution with PSF. The combined use of circular and figure-of-eight coils provides comprehensive imaging coverage. Notably, this research presents a practical method for estimating both circular and figure-of-eight coils' transfer functions, achieving 12.9 dB SNR improvement with high-quality factor coils. A simplified breast model is rotated 16 steps, and sector scan conductive boundary images are reconstructed for both coils. A two-dimensional image of a breast model is obtained by combining images from two different coils. These findings offer significant advancements in MAET-MI imaging, particularly in low SNR environments.

Authors

Institutions

Publication Details

Journal
International Journal for Numerical Methods in Biomedical Engineering
Published
2025-07-01
DOI
https://doi.org/10.1002/cnm.70063
Primary Topic
Electrical and Bioimpedance Tomography
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Enhancing Imaging Performance and Resolution in Magneto‐Acoustic Electrical Tomography With Magnetic Field Measurements ( MAET ‐ MI ) Using Figure‐of‐Eight and High‐Quality Factor Circular Coils

Ahmet Önder Tetik, N.G. Gençer
International Journal for Numerical Methods in Biomedical Engineering
Electrical and Bioimpedance Tomography
article

Enhancing Imaging Performance and Resolution in Magneto‐Acoustic Electrical Tomography With Magnetic Field Measurements ( MAET ‐ MI ) Using Figure‐of‐Eight and High‐Quality Factor Circular Coils

Ahmet Önder Tetik, N.G. Gençer
article en

Abstract

Magneto-acousto-electrical tomography with magnetic field measurement technique (MAET-MI) is a hybrid imaging method that brings high spatial resolution of ultrasound imaging in electrical impedance tomography. This study investigates the impact of the quality factor of circular and figure-of-eight coils on the imaging performance of MAET-MI. Induced MAET signals on the circular coil are accurately obtained by modeling a circuit representation of an air-cored circular coil and deriving its transfer function through impedance measurements. The study demonstrates a significant improvement in signal-to-noise ratio (SNR) using high-quality factor coils compared to unity quality factor coils. Additionally, a 16-element linear phased array (LPA) ultrasound transducer, an air core circular coil, and a figure-of-eight coil are numerically modeled to obtain sector scan images of two-dimensional conductivity distributions. Point spread function (PSF) is characterized, and the lateral resolution of sector scan conductivity images is enhanced through two-dimensional deconvolution with PSF. The combined use of circular and figure-of-eight coils provides comprehensive imaging coverage. Notably, this research presents a practical method for estimating both circular and figure-of-eight coils' transfer functions, achieving 12.9 dB SNR improvement with high-quality factor coils. A simplified breast model is rotated 16 steps, and sector scan conductive boundary images are reconstructed for both coils. A two-dimensional image of a breast model is obtained by combining images from two different coils. These findings offer significant advancements in MAET-MI imaging, particularly in low SNR environments.

International Journal for Numerical Methods in Biomedical EngineeringVol. 41(7)
Middle East Technical University (TR)
Türkiye Bilimsel ve Teknolojik Araştırma Kurumu
Openalex Percentile: Top 85%
Electrical and Bioimpedance Tomography
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.