Agreement Between Lenstar LS900, Pentacam, and Sirius Devices in Terms of Anterior Segment Parameter Measurements in Keratoconic and Healthy Eyes
The three devices demonstrated agreement for ACD and WTW. The Lenstar LS900 and Pentacam HR were compatible in PD. Keratometry values were in agreement between the three devices in healthy eyes. This agreement did not hold in keratoconus. Using these devices interchangeably to evaluate keratometry in keratoconus may yield misleading results. <b>[<i>J Refract Surg</i>. 2025;41(7):e690-e701.]</b>.