Trueness and precision of different intraoral scanners for shade assessment under variable light conditions-a cross-sectional study

To compare the trueness and precision of shade assessment using different intraoral scanners under variable light conditions. To evaluate shade trueness and precision, color measurements (L*, a*, b*) were obtained using visual assessment, a spectrophotometer (SPM) (VITA Easyshade Advanced V), and two intraoral scanners, intra oral scanner-A(IOS-A) (Carestream 3700, Dexis IS ScanFlow) and intra oral scanner-B(IOS-B) (3Shape Trios 3), were used to scan 20 subjects by single assessor under natural daylight and operatory light. The color difference (ΔE) was calculated. Trueness and precision were assessed for each method across the two lighting conditions. Trueness of the test groups was compared by Analysis of Variance (ANOVA) followed by Tukey’s Post hoc Test for pairwise comparisons and precision by Kruskal Wallis test while pairwise comparison was done by Mann Whitney U Test. P values <0.05 was considered as statistically significant. For trueness, significant differences were observed among the visual method, IOS-A, and IOS-B. Tukey’s test revealed a significant difference between the trueness of IOS-A and IOS-B, while no significant differences were detected between the visual method and either scanner. Regarding precision, significant variations were found among the test groups. A significant difference was noted between the visual method and IOS-A, with a highly significant difference between IOS-A and IOS-B. However, no significant differences were observed between SPM and the visual method or IOS-B, nor between the visual method and IOS-B. IOS-B shows better precision for shade determination than IOS-A, while the visual method is more reproducible than IOS-A. No significant differences in trueness and precision were found between the visual method and IOS-B.

Authors

Institutions

Publication Details

Journal
Digital Dentistry Journal
Published
2025-07-05
DOI
https://doi.org/10.1016/j.ddj.2025.100024
Primary Topic
Dental materials and restorations
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Trueness and precision of different intraoral scanners for shade assessment under variable light conditions-a cross-sectional study

Mosin Shaikh, Smita Musani, Taha Attarwala, Aamir Godil et al.
Digital Dentistry Journal
Dental materials and restorations
article

Trueness and precision of different intraoral scanners for shade assessment under variable light conditions-a cross-sectional study

Mosin Shaikh, Smita Musani, Taha Attarwala, Aamir Godil, Rutika Naik, Ahmed Ibrahim Sirkhot, Afnan Sarguroh
article en

Abstract

To compare the trueness and precision of shade assessment using different intraoral scanners under variable light conditions. To evaluate shade trueness and precision, color measurements (L*, a*, b*) were obtained using visual assessment, a spectrophotometer (SPM) (VITA Easyshade Advanced V), and two intraoral scanners, intra oral scanner-A(IOS-A) (Carestream 3700, Dexis IS ScanFlow) and intra oral scanner-B(IOS-B) (3Shape Trios 3), were used to scan 20 subjects by single assessor under natural daylight and operatory light. The color difference (ΔE) was calculated. Trueness and precision were assessed for each method across the two lighting conditions. Trueness of the test groups was compared by Analysis of Variance (ANOVA) followed by Tukey’s Post hoc Test for pairwise comparisons and precision by Kruskal Wallis test while pairwise comparison was done by Mann Whitney U Test. P values <0.05 was considered as statistically significant. For trueness, significant differences were observed among the visual method, IOS-A, and IOS-B. Tukey’s test revealed a significant difference between the trueness of IOS-A and IOS-B, while no significant differences were detected between the visual method and either scanner. Regarding precision, significant variations were found among the test groups. A significant difference was noted between the visual method and IOS-A, with a highly significant difference between IOS-A and IOS-B. However, no significant differences were observed between SPM and the visual method or IOS-B, nor between the visual method and IOS-B. IOS-B shows better precision for shade determination than IOS-A, while the visual method is more reproducible than IOS-A. No significant differences in trueness and precision were found between the visual method and IOS-B.

Digital Dentistry JournalVol. 2(2)
University of Toronto (CA), ACPM Dental College and Hospital (IN)
Climate action
Openalex Percentile: Top 83%
Dental materials and restorations
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.