The Role of Calibration Practices in Optimizing Continuous Glucose Monitor Accuracy in Critically Ill Patients
Our feasibility study demonstrated an improvement in CGM accuracy with POC BG calibrations in ICU patients. Further research is needed to understand optimal implementation strategies and impact on outcomes. CHEST Critical Care 2025; 3(4):100193.
Authors
- Eileen R. Faulds (ORCID: https://orcid.org/0000-0003-2315-1146)
- Kathleen Marie Dungan (ORCID: https://orcid.org/0000-0003-1289-1595)
- Molly McNett (ORCID: https://orcid.org/0000-0003-0222-0421)
- Chyongchiou J. Lin (ORCID: https://orcid.org/0000-0001-9441-3508)
- Brooke Lee (ORCID: https://orcid.org/0000-0001-8795-8470)
- Matthew C. Exline (ORCID: https://orcid.org/0000-0001-6797-158X)
- Melanie Natasha Rayan (ORCID: https://orcid.org/0009-0005-0646-5508)
- Amanie Rasul
- Laureen Jones
Institutions
- The Ohio State University Wexner Medical Center (US)
Publication Details
- Journal
- CHEST Critical Care
- Published
- 2025-07-05
- DOI
- https://doi.org/10.1016/j.chstcc.2025.100193
- Citations
- 1
- Primary Topic
- Hemodynamic Monitoring and Therapy
- Type
- article
- Field-Weighted Citation Impact
- 2.01
Funders
- Dexcom
- Center for Clinical and Translational Science, Ohio State University
- National Institute of Nursing Research
- National Center for Advancing Translational Sciences