Investigation of intracochlear electrical fields with spread of excitation and voltage matrix in cochlear implant users and their link to speech perception
SoE and VM are closely related measurements, however they are not identical. The neural information in SoE can't be ignored and VM can't replace SoE. The VM separation index appeared to be a promising approach for predicting listening performance. Nevertheless, further research is required to corroborate this finding.
Authors
- Pascal Nachtigäller (ORCID: https://orcid.org/0000-0001-7705-0848)
- Tobias Rader (ORCID: https://orcid.org/0000-0001-7250-2833)
- Tobias Weißgerber (ORCID: https://orcid.org/0000-0002-0327-1274)
- Uwe Baumann (ORCID: https://orcid.org/0000-0002-1295-2661)
Institutions
- Goethe University Frankfurt (DE)
- University Hospital Frankfurt (DE)
- Ludwig-Maximilians-Universität München (DE)
Publication Details
- Journal
- Hearing Research
- Published
- 2025-07-05
- DOI
- https://doi.org/10.1016/j.heares.2025.109357
- Citations
- 1
- Primary Topic
- Hearing Loss and Rehabilitation
- Type
- article
- Field-Weighted Citation Impact
- 1.43