Pengaruh Variasi Arus Tabung (mA) Terhadap Kualitas Citra Pada Pesawat Computed Tomography (CT) Scan di RSUD Bali Mandara

Computed Tomography (CT) Scan is one of the essential diagnostic tools in the medical field. CT scan image quality is represented by the Signal-to-Noise Ratio (SNR) and Contrast-to-Noise Ratio (CNR). This study was conducted to understand the effect of tube current variation on image quality in CT scan examinations. The CT scan parameters were set at an X-ray tube voltage of 120 kV, an exposure time of 1 second, and tube current variations of 100 mA, 200 mA, 300 mA, 400 mA, and 500 mA. From 25 image data obtained from scanning the Tos phantom on the CT scan, data were collected on the average signal intensity of the material, the average signal intensity of the background, and the standard deviation of the background signal intensity obtained from the Region of Interest (ROI) readings on each scanned phantom image. The Pearson correlation test results between tube current and SNR values showed that the Pearson correlation coefficient for polypropylene, nylon, acrylic, delrin, and air were 0.993, 0.992, 0.993, 0.992, and 0.992, respectively. The linear regression correlation coefficient (r) for polypropylene, nylon, acrylic, delrin, and air were also 0.993, 0.992, 0.993, 0.992, and 0.992, respectively. Similarly, the Pearson correlation test results between tube current and CNR values showed that the Pearson correlation coefficient for polypropylene, nylon, acrylic, delrin, and air were 0.992, 0.992, 0.993, 0.992, and 0.992, respectively. The linear regression correlation coefficient (r) for polypropylene, nylon, acrylic, delrin, and air were also 0.992, 0.992, 0.993, 0.992, and 0.992, respectively. These results indicate that for all phantom materials, the tube current value has a strong positive correlation with both SNR and CNR. The SNR and CNR values increase linearly with increasing tube current, suggesting that a higher tube current can improve CT scan image quality.

Authors

Institutions

Publication Details

Journal
Kappa Journal
Published
2025-04-27
DOI
https://doi.org/10.29408/kpj.v9i1.29378
Primary Topic
Computer Science and Engineering
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Pengaruh Variasi Arus Tabung (mA) Terhadap Kualitas Citra Pada Pesawat Computed Tomography (CT) Scan di RSUD Bali Mandara

Putu Suardana, Ni Komang Tri Suandayani, Ida Bagus Alit Paramarta, I Nengah Sandi et al.
Kappa Journal
Computer Science and Engineering
article

Pengaruh Variasi Arus Tabung (mA) Terhadap Kualitas Citra Pada Pesawat Computed Tomography (CT) Scan di RSUD Bali Mandara

Putu Suardana, Ni Komang Tri Suandayani, Ida Bagus Alit Paramarta, I Nengah Sandi, Hartono Setiawan, Gusti Ngurah Sutapa, I Dewa Ayu Sintiya Biantari
article en

Abstract

Computed Tomography (CT) Scan is one of the essential diagnostic tools in the medical field. CT scan image quality is represented by the Signal-to-Noise Ratio (SNR) and Contrast-to-Noise Ratio (CNR). This study was conducted to understand the effect of tube current variation on image quality in CT scan examinations. The CT scan parameters were set at an X-ray tube voltage of 120 kV, an exposure time of 1 second, and tube current variations of 100 mA, 200 mA, 300 mA, 400 mA, and 500 mA. From 25 image data obtained from scanning the Tos phantom on the CT scan, data were collected on the average signal intensity of the material, the average signal intensity of the background, and the standard deviation of the background signal intensity obtained from the Region of Interest (ROI) readings on each scanned phantom image. The Pearson correlation test results between tube current and SNR values showed that the Pearson correlation coefficient for polypropylene, nylon, acrylic, delrin, and air were 0.993, 0.992, 0.993, 0.992, and 0.992, respectively. The linear regression correlation coefficient (r) for polypropylene, nylon, acrylic, delrin, and air were also 0.993, 0.992, 0.993, 0.992, and 0.992, respectively. Similarly, the Pearson correlation test results between tube current and CNR values showed that the Pearson correlation coefficient for polypropylene, nylon, acrylic, delrin, and air were 0.992, 0.992, 0.993, 0.992, and 0.992, respectively. The linear regression correlation coefficient (r) for polypropylene, nylon, acrylic, delrin, and air were also 0.992, 0.992, 0.993, 0.992, and 0.992, respectively. These results indicate that for all phantom materials, the tube current value has a strong positive correlation with both SNR and CNR. The SNR and CNR values increase linearly with increasing tube current, suggesting that a higher tube current can improve CT scan image quality.

Kappa JournalVol. 9(1)
Udayana University (ID), Rumah Sakit Umum Pusat Dr. Sardjito (ID)
Openalex Percentile: Top 92%
Computer Science and Engineering
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.