Pengaruh Focus Film Distance (FFD) terhadap Nilai Uji Kolimasi pada Kolimator Pesawat Sinar-X Stasioner untuk Pemeriksaan Pasien

A study has been conducted on the effect of Focus Film Distance (FFD) on the collimation test values of stationary X-ray unit collimators for patient examinations. This research aims to determine the extent to which FFD affects collimation test values. The FFD variations used were 90 cm, 95 cm, 100 cm, 105 cm, 110 cm, 115 cm, and 120 cm. There are three tests in the collimation test: illumination test, collimation field discrepancy with X-ray beam, and X-ray beam perpendicularity. The illumination test was conducted using a lux meter, which was then averaged per FFD. The resulting illumination values corresponding to the FFD variations were 425.25 lux, 381.55 lux, 343.90 lux, 311.80 lux, 283.50 lux, 259.45 lux, and 238.20 lux. These results conform to the inverse square law. For the collimation field discrepancy values |dX| and |dY|, corresponding to the FFD variations, the values were 0.68 and 0.62; 0.32 and 0.34; 0.02 and 0.02; 0.10 and 0.08; 0.18 and 0.20; 0.34 and 0.32; 0.48 and 0.50, respectively. The X-ray beam perpendicularity value remained constant for all FFD variations at 1.03º. The values obtained from the three tests comply with the Indonesian Ministry of Health Regulation No. 1250/MENKES/SK/XII/2009 concerning Quality Control Guidelines for Radiodiagnostic Equipment.

Authors

Institutions

Publication Details

Journal
Kappa Journal
Published
2025-04-24
DOI
https://doi.org/10.29408/kpj.v9i1.28903
Primary Topic
Computer Science and Engineering
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Pengaruh Focus Film Distance (FFD) terhadap Nilai Uji Kolimasi pada Kolimator Pesawat Sinar-X Stasioner untuk Pemeriksaan Pasien

I. G. Antha Kasmawan, Ida Bagus Alit Paramarta, I Wayan Sudarsana, Gusti Ngurah Sutapa et al.
4 links
Kappa Journal
Computer Science and Engineering
article

Pengaruh Focus Film Distance (FFD) terhadap Nilai Uji Kolimasi pada Kolimator Pesawat Sinar-X Stasioner untuk Pemeriksaan Pasien

I. G. Antha Kasmawan, Ida Bagus Alit Paramarta, I Wayan Sudarsana, Gusti Ngurah Sutapa, Ni Nyoman Ratini, Nathaniel Marcellino
article en

Abstract

A study has been conducted on the effect of Focus Film Distance (FFD) on the collimation test values of stationary X-ray unit collimators for patient examinations. This research aims to determine the extent to which FFD affects collimation test values. The FFD variations used were 90 cm, 95 cm, 100 cm, 105 cm, 110 cm, 115 cm, and 120 cm. There are three tests in the collimation test: illumination test, collimation field discrepancy with X-ray beam, and X-ray beam perpendicularity. The illumination test was conducted using a lux meter, which was then averaged per FFD. The resulting illumination values corresponding to the FFD variations were 425.25 lux, 381.55 lux, 343.90 lux, 311.80 lux, 283.50 lux, 259.45 lux, and 238.20 lux. These results conform to the inverse square law. For the collimation field discrepancy values |dX| and |dY|, corresponding to the FFD variations, the values were 0.68 and 0.62; 0.32 and 0.34; 0.02 and 0.02; 0.10 and 0.08; 0.18 and 0.20; 0.34 and 0.32; 0.48 and 0.50, respectively. The X-ray beam perpendicularity value remained constant for all FFD variations at 1.03º. The values obtained from the three tests comply with the Indonesian Ministry of Health Regulation No. 1250/MENKES/SK/XII/2009 concerning Quality Control Guidelines for Radiodiagnostic Equipment.

Kappa JournalVol. 9(1)
Udayana University (ID), Rumah Sakit Umum Pusat Dr. Sardjito (ID)
Universitas Udayana, Tenaga Nasional Berhad, International Atomic Energy Agency, Universitas Sebelas Maret, Universitas Hasanuddin, Universitas Sumatera Utara, Universitas Jember, Fakultas Matematika dan Ilmu Pengetahuan Alam, Universitas Brawijaya
Openalex Percentile: Top 92%
Computer Science and Engineering
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Pengaruh Focus Film Distance (FFD) terhadap Nilai Uji Kolimasi pada Kolimator Pesawat Sinar-X Stasioner untuk Pemeriksaan Pasien — I. G. Antha Kasmawan, Ida Bagus Alit Paramarta, et al. · Kappa Journal (2025) | TGRS Research Map | TGRS