Method for obtaining and dual-energy processing of X-ray sinograms for identifying low-contrast ubstances in CT using sources with a continuous radiation spectrum
The article proposes and implements a method of dual-energy CT. The method allows the contrast-enhanced differentiation of one substance from another when substances have similar chemical compositions and densities. In this case, the use of standard dual-energy CT methods does not allow the unambiguous identification of the desired substance. This is because X-ray sources emit non-monochromatic radiation, and changes in object thickness lead to significant absorption of low-energy radiation. The article justifies the use of a design solution to reduce the influence of variable object thickness on the absorption of low-energy X-ray radiation. This allows dual-energy transformation to be performed not for CT slices but for sinograms obtained at different energy spectra of X-ray radiation. After reconstructing the dual-energy sinogram on a CT slice, the desired substance is unambiguously identified. The successful implementation of this method is demonstrated in the example of beryl particles embedded in muscovite. In CT slices obtained after reconstructing dual-energy sinograms, beryl particles with sizes as small as 1 mm are clearly distinguished with a total sample thickness of 30 mm.
Authors
- A. A. Komarskiy (ORCID: https://orcid.org/0000-0002-1102-8282)
- S. R. Korzhenevskiy (ORCID: https://orcid.org/0000-0003-3669-5230)
- V.V. Krinitsin
Publication Details
- Journal
- Computer Optics
- Published
- 2025-04-01
- DOI
- https://doi.org/10.18287/2412-6179-co-1485
- Primary Topic
- Advanced X-ray and CT Imaging
- Type
- article
- Field-Weighted Citation Impact
- 0.00
Funders
- Russian Science Foundation