Machine Description File-based Monte Carlo Simulation Modeling and its Validation Using Pencil Beam Scanning Proton Therapy Commissioning Data

Purpose: This study aims to detail the machine description file (MDF)-based modeling and validation of the independent Monte Carlo (MC) dose engine using commissioning beam data of a multi-gantry Proteus Plus proton pencil beam scanning (PBS) machine, also, enhancing the understanding of its operational parameters and clinical applications. Materials and Methods: The commissioning of the PBS machine involved measuring key parameters such as integrated depth dose, spot size, and absolute dose using calibrated detectors and ionization chambers. These measurements facilitated the modeling of parameters like energy spread and spot tuning within the RayStation treatment planning system and the tool for particle simulation (TOPAS) Geant4-based MC framework. The validation process employed a MDF containing commissioning beam data, assessing 36 spread-out Bragg peak plans by evaluating range, modulation, distal dose fall-off, and lateral profile characteristics at varying depths. In addition, five patient-specific quality assurance (PSQA) plans were created to evaluate nonhomogeneous fields, following American Association of Physicists in Medicine TG report 185 guidelines. Results: The validation results indicated strong agreement between modeled and measured data, with acceptable comparisons observed between RayStation and TOPAS outputs for homogeneous fields. Nonhomogeneous PSQA results demonstrated that over 95% of cases met the gamma passing criteria of 3%/3 mm. Furthermore, assessments of water equivalent thickness in heterogeneous tissues – such as muscle, fat, bone, and liver – showed discrepancies within 0.5% of the range uncertainty, confirming the accuracy of the modeling approach. Conclusion: The MDF-based modeling and validation process is a robust and practical approach for clinical environments, effectively supporting proton therapy research and ensuring quality assurance in treatment planning. The methodologies developed in this study provide valuable insights and tools for researchers utilizing TOPAS MC as a clinical research instrument, paving the way for advancements in proton therapy techniques and patient-specific treatment plans.

Authors

Institutions

Publication Details

Journal
Journal of Medical Physics
Published
2025-04-01
DOI
https://doi.org/10.4103/jmp.jmp_10_25
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Machine Description File-based Monte Carlo Simulation Modeling and its Validation Using Pencil Beam Scanning Proton Therapy Commissioning Data

G.T. Bholane, Umesh Bharat Gayake, S.D. Dhole, B.J. Patil et al.
1 links
Journal of Medical Physics
Advanced Radiotherapy Techniques
article

Machine Description File-based Monte Carlo Simulation Modeling and its Validation Using Pencil Beam Scanning Proton Therapy Commissioning Data

G.T. Bholane, Umesh Bharat Gayake, S.D. Dhole, B.J. Patil, Siddhartha Laskar, Kantaram Darekar, Lalit Chaudhary
article en

Abstract

Purpose: This study aims to detail the machine description file (MDF)-based modeling and validation of the independent Monte Carlo (MC) dose engine using commissioning beam data of a multi-gantry Proteus Plus proton pencil beam scanning (PBS) machine, also, enhancing the understanding of its operational parameters and clinical applications. Materials and Methods: The commissioning of the PBS machine involved measuring key parameters such as integrated depth dose, spot size, and absolute dose using calibrated detectors and ionization chambers. These measurements facilitated the modeling of parameters like energy spread and spot tuning within the RayStation treatment planning system and the tool for particle simulation (TOPAS) Geant4-based MC framework. The validation process employed a MDF containing commissioning beam data, assessing 36 spread-out Bragg peak plans by evaluating range, modulation, distal dose fall-off, and lateral profile characteristics at varying depths. In addition, five patient-specific quality assurance (PSQA) plans were created to evaluate nonhomogeneous fields, following American Association of Physicists in Medicine TG report 185 guidelines. Results: The validation results indicated strong agreement between modeled and measured data, with acceptable comparisons observed between RayStation and TOPAS outputs for homogeneous fields. Nonhomogeneous PSQA results demonstrated that over 95% of cases met the gamma passing criteria of 3%/3 mm. Furthermore, assessments of water equivalent thickness in heterogeneous tissues – such as muscle, fat, bone, and liver – showed discrepancies within 0.5% of the range uncertainty, confirming the accuracy of the modeling approach. Conclusion: The MDF-based modeling and validation process is a robust and practical approach for clinical environments, effectively supporting proton therapy research and ensuring quality assurance in treatment planning. The methodologies developed in this study provide valuable insights and tools for researchers utilizing TOPAS MC as a clinical research instrument, paving the way for advancements in proton therapy techniques and patient-specific treatment plans.

Journal of Medical PhysicsVol. 50(2)
Abasaheb Garware College (IN), Homi Bhabha National Institute (IN), Tata Memorial Hospital (IN), Advanced Centre for Treatment, Research and Education in Cancer (IN), Savitribai Phule Pune University (IN)
Openalex Percentile: Top 78%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.