Influence of Field of View and Bowtie Filtration on Cone Beam Computed Tomography Image Quality and Scatter-to-Primary Ratio

Purpose/Aim: The image quality (IQ) of cone-beam computed tomography (CBCT) is often reduced due to X-ray scatter, causing issues such as shading, skin-line artifacts, decreased contrast-to-noise ratio, and inaccurate computed tomography (CT) numbers. This study establishes six metrics for assessing IQ, focusing on both traditional metrics, such as contrast-to-noise ratio, and clinically relevant measures of CT signal accuracy. Using a commercial CBCT system for image-guided radiation therapy (IGRT), the study examines how these metrics vary with axial field-of-view (FOV z ) and bowtie filter use to understand the effects of X-ray scatter on IQ. Materials and Methods: Catphan-600 phantom was scanned at five longitudinal FOV z settings (2–27 cm, Superior-Inferior) with and without a bowtie filter, and all software-based scatter corrections were disabled. Six metrics were evaluated: shading (m shading ), periphery accuracy (m periphery ), noise (m noise ), contrast-to-noise ratio (m CNR ), CT number accuracy (m CT# ), and linearity (m linearity ). Results: All six metrics demonstrated a notable decline in IQ as the FOV z increased from 2 to 27 cm. Specifically, the CNR decreased by half, while m shading increased by 250 HU. The bowtie filter improved CT number accuracy at the periphery by approximately 100–140 HU, partially mitigating the impact of a larger FOV z on IQ. Conclusions: As the FOV z increases, quantitative assessments reveal significant artifacts. Using a bowtie filter improves CNR and CT number accuracy while reducing shading and skin-line artifacts. For enhanced IQ in clinical therapy, minimizing the FOV z is recommended. The evaluation framework established in this study provides a valuable tool for system comparison and assessing scatter correction techniques, aiding in accurate low-contrast detection and supporting advancements in online and adaptive radiotherapy.

Authors

Institutions

Publication Details

Journal
Journal of Medical Physics
Published
2025-04-01
DOI
https://doi.org/10.4103/jmp.jmp_191_24
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Influence of Field of View and Bowtie Filtration on Cone Beam Computed Tomography Image Quality and Scatter-to-Primary Ratio

Noor Mail, Khalid M Alshamrani, Behram Khan
Journal of Medical Physics
Medical Imaging Techniques and Applications
article

Influence of Field of View and Bowtie Filtration on Cone Beam Computed Tomography Image Quality and Scatter-to-Primary Ratio

Noor Mail, Khalid M Alshamrani, Behram Khan
article en

Abstract

Purpose/Aim: The image quality (IQ) of cone-beam computed tomography (CBCT) is often reduced due to X-ray scatter, causing issues such as shading, skin-line artifacts, decreased contrast-to-noise ratio, and inaccurate computed tomography (CT) numbers. This study establishes six metrics for assessing IQ, focusing on both traditional metrics, such as contrast-to-noise ratio, and clinically relevant measures of CT signal accuracy. Using a commercial CBCT system for image-guided radiation therapy (IGRT), the study examines how these metrics vary with axial field-of-view (FOV z ) and bowtie filter use to understand the effects of X-ray scatter on IQ. Materials and Methods: Catphan-600 phantom was scanned at five longitudinal FOV z settings (2–27 cm, Superior-Inferior) with and without a bowtie filter, and all software-based scatter corrections were disabled. Six metrics were evaluated: shading (m shading ), periphery accuracy (m periphery ), noise (m noise ), contrast-to-noise ratio (m CNR ), CT number accuracy (m CT# ), and linearity (m linearity ). Results: All six metrics demonstrated a notable decline in IQ as the FOV z increased from 2 to 27 cm. Specifically, the CNR decreased by half, while m shading increased by 250 HU. The bowtie filter improved CT number accuracy at the periphery by approximately 100–140 HU, partially mitigating the impact of a larger FOV z on IQ. Conclusions: As the FOV z increases, quantitative assessments reveal significant artifacts. Using a bowtie filter improves CNR and CT number accuracy while reducing shading and skin-line artifacts. For enhanced IQ in clinical therapy, minimizing the FOV z is recommended. The evaluation framework established in this study provides a valuable tool for system comparison and assessing scatter correction techniques, aiding in accurate low-contrast detection and supporting advancements in online and adaptive radiotherapy.

Journal of Medical PhysicsVol. 50(2)
King Saud bin Abdulaziz University for Health Sciences (SA), University of Toronto (CA), King Abdullah International Medical Research Center (SA), UPMC Hillman Cancer Center (US), University of Pittsburgh Medical Center (US), National Guard Health Affairs (SA)
Openalex Percentile: Top 76%
Medical Imaging Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.