Investigation of Spot Size Variations in Multiroom Proton Therapy Systems and Their Clinical Significance: An In silico Study

Background and Purpose: In multiroom proton therapy facilities, maintaining beam consistency is critical, particularly when treatment interruptions occur due to machine downtime. Among beam parameters, spot size frequently exceeds tolerance limits, potentially compromising treatment accuracy. This study aims to assess the clinical implications of spot size variations and establish benchmarks for beam matching, with a specific focus on the spot size parameter in proton therapy using tool for particle simulation (TOPAS), Monte Carlo (MC) simulation. Materials and Methods: The study analyzed the effects of spot size deviations (±0.3 mm and ± 0.6 mm) on proton therapy treatment plans using TOPAS MC simulations. The five variable spot sizes models were created in the RayStation treatment planning system by simulating the known spot size shift error in the TOPAS for 33 proton energies ranging from 70.18 to 226.2 MeV. These models were evaluated using both homogeneous phantom fields and heterogeneous clinical fields targeting the pelvis, brain, and prostate. Key dosimetric metrics target, coverage (TC), conformity index, and homogeneity index, were assessed. In addition, two-dimensional gamma analysis was performed at tolerances of 1%/1 mm, 2%/2 mm, and 3%/2 mm to quantify the clinical impact of spot size variations on treatment delivery accuracy. Results: The evaluation indicated that Model −0.6 mm achieved the highest target coverage (TC), while Model 0.6 mm resulted in the lowest. Comparative analysis within the homogeneous phantom revealed marked variations in TC and conformity indices among the tested models. Clinical investigation revealed that TC in the pelvic area was constant, but that TC in the brain and prostate was more sensitive to changes in spot size. Gamma analysis showed superior passing rates for the model ± 0.3 mm, particularly at 2%/2 mm and 3%/2 mm criteria, confirming its suitability for optimal treatment accuracy. Conclusion: Spot size variations significantly influence the accuracy of proton therapy, with deviations of ± 0.3 mm yielding the most favorable results. Adhering to this tolerance ensures consistent beam matching and precise treatment delivery across various clinical sites, supporting the reliability of multiroom proton therapy systems.

Authors

Institutions

Publication Details

Journal
Journal of Medical Physics
Published
2025-04-01
DOI
https://doi.org/10.4103/jmp.jmp_8_25
Primary Topic
Radiation Therapy and Dosimetry
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Investigation of Spot Size Variations in Multiroom Proton Therapy Systems and Their Clinical Significance: An In silico Study

Umesh Bharat Gayake, S.D. Dhole, Siddhartha Laskar, B.J. Patil et al.
Journal of Medical Physics
Radiation Therapy and Dosimetry
article

Investigation of Spot Size Variations in Multiroom Proton Therapy Systems and Their Clinical Significance: An In silico Study

Umesh Bharat Gayake, S.D. Dhole, Siddhartha Laskar, B.J. Patil, Kantaram Darekar, Lalit Chaudhary
article en

Abstract

Background and Purpose: In multiroom proton therapy facilities, maintaining beam consistency is critical, particularly when treatment interruptions occur due to machine downtime. Among beam parameters, spot size frequently exceeds tolerance limits, potentially compromising treatment accuracy. This study aims to assess the clinical implications of spot size variations and establish benchmarks for beam matching, with a specific focus on the spot size parameter in proton therapy using tool for particle simulation (TOPAS), Monte Carlo (MC) simulation. Materials and Methods: The study analyzed the effects of spot size deviations (±0.3 mm and ± 0.6 mm) on proton therapy treatment plans using TOPAS MC simulations. The five variable spot sizes models were created in the RayStation treatment planning system by simulating the known spot size shift error in the TOPAS for 33 proton energies ranging from 70.18 to 226.2 MeV. These models were evaluated using both homogeneous phantom fields and heterogeneous clinical fields targeting the pelvis, brain, and prostate. Key dosimetric metrics target, coverage (TC), conformity index, and homogeneity index, were assessed. In addition, two-dimensional gamma analysis was performed at tolerances of 1%/1 mm, 2%/2 mm, and 3%/2 mm to quantify the clinical impact of spot size variations on treatment delivery accuracy. Results: The evaluation indicated that Model −0.6 mm achieved the highest target coverage (TC), while Model 0.6 mm resulted in the lowest. Comparative analysis within the homogeneous phantom revealed marked variations in TC and conformity indices among the tested models. Clinical investigation revealed that TC in the pelvic area was constant, but that TC in the brain and prostate was more sensitive to changes in spot size. Gamma analysis showed superior passing rates for the model ± 0.3 mm, particularly at 2%/2 mm and 3%/2 mm criteria, confirming its suitability for optimal treatment accuracy. Conclusion: Spot size variations significantly influence the accuracy of proton therapy, with deviations of ± 0.3 mm yielding the most favorable results. Adhering to this tolerance ensures consistent beam matching and precise treatment delivery across various clinical sites, supporting the reliability of multiroom proton therapy systems.

Journal of Medical PhysicsVol. 50(2)
Abasaheb Garware College (IN), Homi Bhabha National Institute (IN), Tata Memorial Hospital (IN), Advanced Centre for Treatment, Research and Education in Cancer (IN), Savitribai Phule Pune University (IN)
Openalex Percentile: Top 74%
Radiation Therapy and Dosimetry
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.