VALIDATING INTERNAL DENSITY CALIBRATION IN THE PROXIMAL HUMERUS TO ESTIMATE BONE STIFFNESS FOR STEMLESS SHOULDER ARTHROPLASTY

Stemless humeral head components have emerged as a popular choice for patients undergoing shoulder arthroplasty for end-stage OA since they preserve non-diseased bone for future surgical revisions. Current pre-operative clinical measures are limited in assessing volumetric bone mineral density (vBMD) and mechanical properties in the region of bone directly supporting the component. Gold-standard phantom calibration, used to determine vBMD in CT images, is seldom utilized in clinical practice requiring alternative density measures for accurate vBMD. Internal density calibration using internal tissues as references has yet to be validated in the proximal humerus, and vBMD values have yet to be linked to finite element model (FEM) estimated stiffness in the context of stemless shoulder arthroplasty. 1) To determine the correlation between vBMD and finite element model (FEM) estimated stiffness 2) To determine the bias in internal density-based vBMD using three different referent tissue combinations compared to phantom-based vBMD, in the proximal humerus. Non-pathologic cadaveric single-energy CT images (n = 25), containing a K 2 HPO 4 phantom, were used to analyze a 10 mm region directly below the anatomic neck. Phantom-based vBMD was calculated for each region and used as input to image-based FEMs (ROD). Internal calibration used air (A), adipose (A), skeletal muscle (M), and cortical bone (C) to generate calibrated images from three different referent tissue combinations (AACM, ACM, AAC). Images were used to generate FEMs for each tissue combination. Results were compared between vBMD (mg K 2 HPO 4 /cc) and apparent modulus (E app ) for each internal calibration tissue combination to the phantom calibration using linear regression. Bland-Altman analysis was used to determine the agreement between tissue combination and phantom calibration for estimated stiffness values (E app ). Linear regression (Figure 1) showed strong correlations between estimated stiffness and vBMD values for each calibration method (AACM R 2 = 0.7524; ACM R 2 = 0.7723; AAC R 2 = 0.7384; ROD R 2 = 0.7854) and slopes not significantly different from 1 (p < 0.001). Bland-Altman analysis (Figure 2) revealed the ACM tissue combination had the lowest error bounds in apparent modulus, compared to phantom-vBMD derived FEMs, with a mean bias of 80.15 MPa and 95% limits of agreement ranging from -164.55 to 324.86 MPa. The results of this study support the use of internal density calibration as a valid method for using internal density calibrated images as input to FEMs for estimating stiffness in the proximal humerus. The ACM tissue combination provided the highest agreement with the gold standard phantom calibration. This internal density calibration method may provide a solution for determining vBMD in patients undergoing shoulder arthroplasty for end-stage OA where phantoms are not present in the CT image. By linking bone density measures with estimated stiffness values, the mechanical properties of bone in the region supporting the humeral component are considered, which has the potential to improve preoperative planning for stemless shoulder arthroplasty. Next steps are to apply the ACM internal calibration method and estimated stiffness values in retrospective CT images from patients who have undergone shoulder arthroplasty for end-stage OA (n = 88) to link surgical outcomes to stiffness measures.

Authors

Institutions

Publication Details

Journal
Osteoarthritis Imaging
Published
2025-01-01
DOI
https://doi.org/10.1016/j.ostima.2025.100321
Primary Topic
Shoulder Injury and Treatment
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

VALIDATING INTERNAL DENSITY CALIBRATION IN THE PROXIMAL HUMERUS TO ESTIMATE BONE STIFFNESS FOR STEMLESS SHOULDER ARTHROPLASTY

Clark R. Dickerson, Jack P. Callaghan, Steven K. Boyd, Nikolas K. Knowles et al.
Osteoarthritis Imaging
Shoulder Injury and Treatment
article

VALIDATING INTERNAL DENSITY CALIBRATION IN THE PROXIMAL HUMERUS TO ESTIMATE BONE STIFFNESS FOR STEMLESS SHOULDER ARTHROPLASTY

Clark R. Dickerson, Jack P. Callaghan, Steven K. Boyd, Nikolas K. Knowles, Chloe Stiles, B. Matheson, G.S. Arthwal
article en

Abstract

Stemless humeral head components have emerged as a popular choice for patients undergoing shoulder arthroplasty for end-stage OA since they preserve non-diseased bone for future surgical revisions. Current pre-operative clinical measures are limited in assessing volumetric bone mineral density (vBMD) and mechanical properties in the region of bone directly supporting the component. Gold-standard phantom calibration, used to determine vBMD in CT images, is seldom utilized in clinical practice requiring alternative density measures for accurate vBMD. Internal density calibration using internal tissues as references has yet to be validated in the proximal humerus, and vBMD values have yet to be linked to finite element model (FEM) estimated stiffness in the context of stemless shoulder arthroplasty. 1) To determine the correlation between vBMD and finite element model (FEM) estimated stiffness 2) To determine the bias in internal density-based vBMD using three different referent tissue combinations compared to phantom-based vBMD, in the proximal humerus. Non-pathologic cadaveric single-energy CT images (n = 25), containing a K 2 HPO 4 phantom, were used to analyze a 10 mm region directly below the anatomic neck. Phantom-based vBMD was calculated for each region and used as input to image-based FEMs (ROD). Internal calibration used air (A), adipose (A), skeletal muscle (M), and cortical bone (C) to generate calibrated images from three different referent tissue combinations (AACM, ACM, AAC). Images were used to generate FEMs for each tissue combination. Results were compared between vBMD (mg K 2 HPO 4 /cc) and apparent modulus (E app ) for each internal calibration tissue combination to the phantom calibration using linear regression. Bland-Altman analysis was used to determine the agreement between tissue combination and phantom calibration for estimated stiffness values (E app ). Linear regression (Figure 1) showed strong correlations between estimated stiffness and vBMD values for each calibration method (AACM R 2 = 0.7524; ACM R 2 = 0.7723; AAC R 2 = 0.7384; ROD R 2 = 0.7854) and slopes not significantly different from 1 (p < 0.001). Bland-Altman analysis (Figure 2) revealed the ACM tissue combination had the lowest error bounds in apparent modulus, compared to phantom-vBMD derived FEMs, with a mean bias of 80.15 MPa and 95% limits of agreement ranging from -164.55 to 324.86 MPa. The results of this study support the use of internal density calibration as a valid method for using internal density calibrated images as input to FEMs for estimating stiffness in the proximal humerus. The ACM tissue combination provided the highest agreement with the gold standard phantom calibration. This internal density calibration method may provide a solution for determining vBMD in patients undergoing shoulder arthroplasty for end-stage OA where phantoms are not present in the CT image. By linking bone density measures with estimated stiffness values, the mechanical properties of bone in the region supporting the humeral component are considered, which has the potential to improve preoperative planning for stemless shoulder arthroplasty. Next steps are to apply the ACM internal calibration method and estimated stiffness values in retrospective CT images from patients who have undergone shoulder arthroplasty for end-stage OA (n = 88) to link surgical outcomes to stiffness measures.

Osteoarthritis ImagingVol. 5
University of Waterloo (CA), University of Calgary (CA), Hand and Upper Limb Clinic (CA), Alberta Bone and Joint Health Institute (CA)
Openalex Percentile: Top 73%
Shoulder Injury and Treatment
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.