Scanning Technologies Adoption for Transtibial Residual Limb Shape Characterization

In prosthetics, 3D scanning technology plays a crucial role in gathering data for designing prosthetic devices. New tools for capturing the external shape of human body parts are rapidly emerging. This study conducted a comparative analysis to assess the accuracy and precision of two handheld 3D scanners, including the Peel 3 3D scanner and a smartphone application, in scanning the residual limb of a transtibial participant. The residual limb was manually measured, and six scans were taken with each scanning system, compared to the reference computed tomography (CT) model. A repetition test was performed to determine the mean value and standard deviation of the scanned models in VXelements software for comparison purposes. The results showed a standard deviation difference between the Peel 3 and CT scans ranging from ± 0.828 to 0.907mm, and from ± 0.823 to ± 0.877mm for the smartphone scans. The repetition test indicated standard deviations ranging from ± 0.531 to 0.599 mm for the Peel 3 and ± 0.458 to 0.690 mm for the smartphone. This analysis evaluated the accuracy of the 3D scanners and identified the essential specifications for their effective use in prosthetics.

Authors

Publication Details

Journal
Journal of Engineering and Sustainable Development
Published
2025-07-01
DOI
https://doi.org/10.31272/jeasd.2818
Primary Topic
Industrial Vision Systems and Defect Detection
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Scanning Technologies Adoption for Transtibial Residual Limb Shape Characterization

Abdulkader Kadauw, Sarah Duraid AlQaissi, Ahmed A.A. AlDuroobi
4 links
Journal of Engineering and Sustainable Development
Industrial Vision Systems and Defect Detection
article

Scanning Technologies Adoption for Transtibial Residual Limb Shape Characterization

Abdulkader Kadauw, Sarah Duraid AlQaissi, Ahmed A.A. AlDuroobi
article en

Abstract

In prosthetics, 3D scanning technology plays a crucial role in gathering data for designing prosthetic devices. New tools for capturing the external shape of human body parts are rapidly emerging. This study conducted a comparative analysis to assess the accuracy and precision of two handheld 3D scanners, including the Peel 3 3D scanner and a smartphone application, in scanning the residual limb of a transtibial participant. The residual limb was manually measured, and six scans were taken with each scanning system, compared to the reference computed tomography (CT) model. A repetition test was performed to determine the mean value and standard deviation of the scanned models in VXelements software for comparison purposes. The results showed a standard deviation difference between the Peel 3 and CT scans ranging from ± 0.828 to 0.907mm, and from ± 0.823 to ± 0.877mm for the smartphone scans. The repetition test indicated standard deviations ranging from ± 0.531 to 0.599 mm for the Peel 3 and ± 0.458 to 0.690 mm for the smartphone. This analysis evaluated the accuracy of the 3D scanners and identified the essential specifications for their effective use in prosthetics.

Journal of Engineering and Sustainable DevelopmentVol. 29(4)
Openalex Percentile: Top 78%
Industrial Vision Systems and Defect Detection
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Scanning Technologies Adoption for Transtibial Residual Limb Shape Characterization — Abdulkader Kadauw, Sarah Duraid AlQaissi, et al. · Journal of Engineering and Sustainable Development (2025) | TGRS Research Map | TGRS