A Rare Hemoglobin Variant Detected for the First Time in Türkiye (Hb Iraq-Halabja): Evaluation of the Effect of Variant Hemoglobins on HbA1c Methods

The commonly used methods for HbA1c measurement are cation-exchange high performance chromatography (CE-HPLC), immunologic method, capillary electrophoresis and boronate affinity HPLC. Hb-variants can reduce the reliability of HbA1c measurements. We aimed to emphasize the importance of step-by-step solutions to the difficulties encountered in HbA1c measurement methods due to Hb-variants. We also aimed to evaluate the advantages and disadvantages of different methods used in HbA1c analysis with the example of the Hb-Iraq-Halabja variant detected for the first time in Türkiye. HbA1c level could not be measured by CE-HPLC (Adams HA-8180V analyzer, Arkray, Japan) and a peak-tailing signal was detected indicating an abnormality between stable HbA1c and HbA0 peaks (concurrent glucose level was 8.55mmol/L). In the second step, HbA1c was able to be measured by immunologic method and boronate affinity method and were found to be 5.87% and 5.90%, respectively (estimated average glucose equivalent 6.66mmol/L). In the last step, genetic analysis was performed due to suspicion of Hb variant and the very rare Hb Iraq-Halabja variant was detected. After genetic verification, HbA1c test was repeated with a different CE-HPLC (HLC-723 G11, T osoh, Tokyo, Japan) and a peak indicating variant Hb was detected between stable HbA1c and HbA0. A remarkable finding was that, unlike the previous CE-HPLC (Arkray) result, HbA1c could be measured as 3.20%. In the concurrent measurement performed on the boronate affinity HPLC (Premier Hb9210, Trinity Biotech, County Wicklow, Ireland), HbA1c result was found to be 5.40% (concurrent glucose 5.22 mmol/L). In addition, concurrent fructosamine serum value was found to be 210 μmol/L (estimated mean glucose equivalent 4.88 mmol/L). The patient's laboratory tests were generally within normal limits, and iron deficiency, hemolytic anemia, and B12-folate deficiency were excluded. The glucose bounding area of hemoglobin is generally preserved and is not affected by common Hb-variants. Boronate affinity and immunologic method (these two methods target glucose bounding areas) that give HbA1c results consistent with the patient's fasting blood glucose and fructosamine results. However, the CE-HPLC method has been observed to either fail to measure HbA1c or to measure falsely low HbA1c due to overlapping peaks of Hb variants.

Authors

Institutions

Publication Details

Journal
Hemoglobin
Published
2025-07-01
DOI
https://doi.org/10.1080/03630269.2025.2524437
Primary Topic
Hemoglobinopathies and Related Disorders
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A Rare Hemoglobin Variant Detected for the First Time in Türkiye (Hb Iraq-Halabja): Evaluation of the Effect of Variant Hemoglobins on HbA1c Methods

Said Sami Erdem, Müşerref Başdemirci, Tevfik Balcı, Öznur Köylü et al.
Hemoglobin
Hemoglobinopathies and Related Disorders
article

A Rare Hemoglobin Variant Detected for the First Time in Türkiye (Hb Iraq-Halabja): Evaluation of the Effect of Variant Hemoglobins on HbA1c Methods

Said Sami Erdem, Müşerref Başdemirci, Tevfik Balcı, Öznur Köylü, Emre Akkaya, Beyza Ünlü
article en

Abstract

The commonly used methods for HbA1c measurement are cation-exchange high performance chromatography (CE-HPLC), immunologic method, capillary electrophoresis and boronate affinity HPLC. Hb-variants can reduce the reliability of HbA1c measurements. We aimed to emphasize the importance of step-by-step solutions to the difficulties encountered in HbA1c measurement methods due to Hb-variants. We also aimed to evaluate the advantages and disadvantages of different methods used in HbA1c analysis with the example of the Hb-Iraq-Halabja variant detected for the first time in Türkiye. HbA1c level could not be measured by CE-HPLC (Adams HA-8180V analyzer, Arkray, Japan) and a peak-tailing signal was detected indicating an abnormality between stable HbA1c and HbA0 peaks (concurrent glucose level was 8.55mmol/L). In the second step, HbA1c was able to be measured by immunologic method and boronate affinity method and were found to be 5.87% and 5.90%, respectively (estimated average glucose equivalent 6.66mmol/L). In the last step, genetic analysis was performed due to suspicion of Hb variant and the very rare Hb Iraq-Halabja variant was detected. After genetic verification, HbA1c test was repeated with a different CE-HPLC (HLC-723 G11, T osoh, Tokyo, Japan) and a peak indicating variant Hb was detected between stable HbA1c and HbA0. A remarkable finding was that, unlike the previous CE-HPLC (Arkray) result, HbA1c could be measured as 3.20%. In the concurrent measurement performed on the boronate affinity HPLC (Premier Hb9210, Trinity Biotech, County Wicklow, Ireland), HbA1c result was found to be 5.40% (concurrent glucose 5.22 mmol/L). In addition, concurrent fructosamine serum value was found to be 210 μmol/L (estimated mean glucose equivalent 4.88 mmol/L). The patient's laboratory tests were generally within normal limits, and iron deficiency, hemolytic anemia, and B12-folate deficiency were excluded. The glucose bounding area of hemoglobin is generally preserved and is not affected by common Hb-variants. Boronate affinity and immunologic method (these two methods target glucose bounding areas) that give HbA1c results consistent with the patient's fasting blood glucose and fructosamine results. However, the CE-HPLC method has been observed to either fail to measure HbA1c or to measure falsely low HbA1c due to overlapping peaks of Hb variants.

HemoglobinVol. 49(4)
City Hospital (GB), Konya Eğitim ve Araştırma Hastanesi (TR)
Openalex Percentile: Top 78%
Hemoglobinopathies and Related Disorders
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.