Impact of magnetic fields from tablets, laptops, smartphones, and household/leisure magnets on cardiac implantable electronic devices
Information processing devices and household/leisure magnets can affect CIEDs when placed in close proximity. Among the devices tested, magnet mode activation did not occur at distances of ≥20 mm. Considering the increasing prevalence of information processing devices and the growing adoption of nonthoracic CIED placements, raising awareness among patients about potential interactions is crucial.
Authors
- Hiroyoshi Miura
- Norio Kamitani
- Ryuta Henmi
- Satoko Tomida
- Aya Miyazaki (ORCID: https://orcid.org/0000-0003-3401-9378)
- Naoto Ohira
- Ryo Sugiura (ORCID: https://orcid.org/0000-0003-4156-8025)
- Keita Shimizu
- Hiroshi Masui
- S Tominaga
- K. Kondo
- Daishi Koyama
Institutions
- Seirei Hamamatsu General Hospital (JP)
Publication Details
- Journal
- Journal of Arrhythmia
- Published
- 2025-06-30
- DOI
- https://doi.org/10.1002/joa3.70106
- Primary Topic
- Cardiac pacing and defibrillation studies
- Type
- article
- Field-Weighted Citation Impact
- 0.00