Comparison of partial discharge (PD) calibrators from 0.1 pC to 10 nC

Main text High-voltage partial discharge (PD) tests require calibration of PD measurement systems using PD calibrators which provide known values of the apparent charge. A set of travelling reference calibrators was circulated among 14 participating laboratories and each laboratory reported apparent change and rise time measured using their own procedures. Comparison reference values (CRV) were calculated for each nominal charge level and calibrator and reasonable agreement of apparent charge calibration was reached. However, agreement of rise time measurements was unexpectedly poor. The scatter in reported rise time values is not due to the instability of the travelling reference and is largely dependent on the implemented methodology. Due to the large range of variables possibly influencing reported results for rise time, the CRV for rise time is omitted. To reach the main text of this paper, click on Final Report . Note that this text is that which appears in Appendix B of the BIPM key comparison database https://www.bipm.org/kcdb/ . The final report has been peer-reviewed and approved for publication by the CCEM, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).

Authors

Institutions

Publication Details

Journal
Metrologia
Published
2025-01-01
DOI
https://doi.org/10.1088/0026-1394/62/1a/01002
Citations
1
Primary Topic
High voltage insulation and dielectric phenomena
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparison of partial discharge (PD) calibrators from 0.1 pC to 10 nC

Jan Hlávacĕk, Wei Yan, G. Crotti, Grzegorz Sadkowski et al.
1 citations
Metrologia
High voltage insulation and dielectric phenomena
article

Comparison of partial discharge (PD) calibrators from 0.1 pC to 10 nC

Jan Hlávacĕk, Wei Yan, G. Crotti, Grzegorz Sadkowski, Ahmet Merev, N Liu, Joni Klüss, Jussi Havunen, Johann Meisner, Jari Hällström, Yazid Hadjadj, M. Zeier, M Berginc, T Aguado, R Vanconcellos
article en
1 citations

Abstract

Main text High-voltage partial discharge (PD) tests require calibration of PD measurement systems using PD calibrators which provide known values of the apparent charge. A set of travelling reference calibrators was circulated among 14 participating laboratories and each laboratory reported apparent change and rise time measured using their own procedures. Comparison reference values (CRV) were calculated for each nominal charge level and calibrator and reasonable agreement of apparent charge calibration was reached. However, agreement of rise time measurements was unexpectedly poor. The scatter in reported rise time values is not due to the instability of the travelling reference and is largely dependent on the implemented methodology. Due to the large range of variables possibly influencing reported results for rise time, the CRV for rise time is omitted. To reach the main text of this paper, click on Final Report . Note that this text is that which appears in Appendix B of the BIPM key comparison database https://www.bipm.org/kcdb/ . The final report has been peer-reviewed and approved for publication by the CCEM, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).

MetrologiaVol. 62(1A)
Physikalisch-Technische Bundesanstalt (DE), National Measurement Institute (AU), RISE Research Institutes of Sweden (SE), IMDEA Materials (ES), Swiss Federal Institute of Metrology (CH), Laboratoire National de Métrologie et d'Essais (FR), Slovenian Institute of Quality and Metrology (SI), Tübitak National Metrology Institute (TR), Český metrologický institut (CZ), Istituto Nazionale di Ricerca Metrologica (IT), Central Office of Measures (PL), National Research Council Canada (CA), National Institute of Metrology (CN), VTT Technical Research Centre of Finland (FI)
Openalex Percentile: Top 89%
High voltage insulation and dielectric phenomena
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.