Key comparison BIPM.RI(I)-K4 of the absorbed dose to water standards in <sup>60</sup>Co gamma radiation beams of the NIST, USA and of the BIPM

Main text A new key comparison of the standards for absorbed dose to water of the National Institute of Standards and Technology (NIST), USA, and of the Bureau International des Poids et Mesures (BIPM) was carried out in the 60 Co radiation beam of the BIPM in October 2023. The comparison result, based on the calibration coefficients for two transfer standards and evaluated as a ratio of the NIST and the BIPM standards for absorbed dose to water, is 1.0019 with a combined standard uncertainty of 5.7 parts in 10 3 . The results are analysed and presented in terms of degrees of equivalence, suitable for entry in the BIPM key comparison database. To reach the main text of this paper, click on Final Report . Note that this text is that which appears in Appendix B of the BIPM key comparison database https://www.bipm.org/kcdb/ . The final report has been peer-reviewed and approved for publication by the CCRI, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).

Authors

Institutions

Publication Details

Journal
Metrologia
Published
2025-01-01
DOI
https://doi.org/10.1088/0026-1394/62/1a/06006
Primary Topic
Radioactivity and Radon Measurements
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Key comparison BIPM.RI(I)-K4 of the absorbed dose to water standards in 60Co gamma radiation beams of the NIST, USA and of the BIPM

R Tosh, C Kessler, Ronaldo Minniti
Metrologia
Radioactivity and Radon Measurements
article

Key comparison BIPM.RI(I)-K4 of the absorbed dose to water standards in 60Co gamma radiation beams of the NIST, USA and of the BIPM

R Tosh, C Kessler, Ronaldo Minniti
article en

Abstract

Main text A new key comparison of the standards for absorbed dose to water of the National Institute of Standards and Technology (NIST), USA, and of the Bureau International des Poids et Mesures (BIPM) was carried out in the 60 Co radiation beam of the BIPM in October 2023. The comparison result, based on the calibration coefficients for two transfer standards and evaluated as a ratio of the NIST and the BIPM standards for absorbed dose to water, is 1.0019 with a combined standard uncertainty of 5.7 parts in 10 3 . The results are analysed and presented in terms of degrees of equivalence, suitable for entry in the BIPM key comparison database. To reach the main text of this paper, click on Final Report . Note that this text is that which appears in Appendix B of the BIPM key comparison database https://www.bipm.org/kcdb/ . The final report has been peer-reviewed and approved for publication by the CCRI, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).

MetrologiaVol. 62(1A)
National Institute of Standards and Technology (US), Bureau international des poids et mesures (FR)
Openalex Percentile: Top 78%
Radioactivity and Radon Measurements
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.