Experimental analysis of blood pressure estimation using electrocardiography and photoplethysmography signals from fingertip measurements

This study investigates the indirect estimation of systolic and diastolic blood pressure using electrocardiography and photoplethysmography signals acquired from the fingertip. Two distinct approaches for blood pressure calculation were evaluated: one based on linear and non-linear models, and another utilizing mathematical models to determine diastolic pressure through mean arterial pressure. The research was conducted using the MAX86150 Evaluation System, with algorithm development and analysis performed in MATLAB. Blood pressure calculations were applied to two patients of different ages, and the results were compared. Additionally, the study examined measurement uncertainty and the role of statistical methods in data validation. It was established that the expanded measurement uncertainty does not exceed ± 11.55 mmHg for systolic blood pressure and ± 7.84 mmHg for diastolic blood pressure at a 95 % confidence level, which complies with the clinical accuracy requirements of the British Hypertension Society and the Association for the Advancement of Medical Instrumentation (AAMI/ANSI/ISO 81060-2:2018). The results showed that the approach based on the mathematical model of mean diastolic pressure demonstrated higher accuracy of arterial pressure estimation. The proposed algorithm exhibited strong reliability, highlighting the need for further validation against reference measurement tools in future research. However, a key limitation of the study is the small sample size — measurements were conducted on only two patients, which significantly restricts the generalizability of the findings.

Authors

Institutions

Publication Details

Journal
ACTA IMEKO
Published
2025-06-20
DOI
https://doi.org/10.21014/actaimeko.v14i2.2096
Citations
1
Primary Topic
Non-Invasive Vital Sign Monitoring
Type
article
Field-Weighted Citation Impact
0.67
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Experimental analysis of blood pressure estimation using electrocardiography and photoplethysmography signals from fingertip measurements

Emanuel Popovici, Maksym Yakovlev, Oleksandr Vasilevskyi, Roman Trishch et al.
1 citations
ACTA IMEKO
Non-Invasive Vital Sign Monitoring
0.67
article

Experimental analysis of blood pressure estimation using electrocardiography and photoplethysmography signals from fingertip measurements

Emanuel Popovici, Maksym Yakovlev, Oleksandr Vasilevskyi, Roman Trishch, Volodymyr Sarana
article en
1 citations

Abstract

This study investigates the indirect estimation of systolic and diastolic blood pressure using electrocardiography and photoplethysmography signals acquired from the fingertip. Two distinct approaches for blood pressure calculation were evaluated: one based on linear and non-linear models, and another utilizing mathematical models to determine diastolic pressure through mean arterial pressure. The research was conducted using the MAX86150 Evaluation System, with algorithm development and analysis performed in MATLAB. Blood pressure calculations were applied to two patients of different ages, and the results were compared. Additionally, the study examined measurement uncertainty and the role of statistical methods in data validation. It was established that the expanded measurement uncertainty does not exceed ± 11.55 mmHg for systolic blood pressure and ± 7.84 mmHg for diastolic blood pressure at a 95 % confidence level, which complies with the clinical accuracy requirements of the British Hypertension Society and the Association for the Advancement of Medical Instrumentation (AAMI/ANSI/ISO 81060-2:2018). The results showed that the approach based on the mathematical model of mean diastolic pressure demonstrated higher accuracy of arterial pressure estimation. The proposed algorithm exhibited strong reliability, highlighting the need for further validation against reference measurement tools in future research. However, a key limitation of the study is the small sample size — measurements were conducted on only two patients, which significantly restricts the generalizability of the findings.

ACTA IMEKOVol. 14(2)
National Aerospace University – Kharkiv Aviation Institute (UA), University College Cork (IE), Central Scientific Research Institute of Armament and Military Equipment of the Armed Forces of Ukraine (UA), The University of Texas at Austin (US)
Openalex Percentile: Top 30%
Non-Invasive Vital Sign Monitoring
0.67
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.