Impact of Portable Radiometers on Irradiance Measurements of LED Photocuring Units

Purpose: The aim of this in vitro study was to evaluate the influence of different models of commercially available portable dental radiometers on the measurement of light irradiance emitted by light-emitting diode (LED) photocuring units. Materials and Methods: Eight LED photocuring units, all emitting light in a single-wavelength spectrum, were tested. Light irradiance (mW/cm2) was measured using six portable dental radiometers: four digital models (D1–D4) and two analog models (A1, A2). Digital model D1 was used as the reference (control). All measurements were conducted under standardized conditions, and each LED–radiometer combination was tested in triplicate. Data were analyzed using Sigma Plot 12.0 (Palo Alto, CA, USA) to verify the assumptions of normality and homogeneity of variances. A one-way analysis of variance (ANOVA) was used to assess the effect of the radiometer model on irradiance values, followed by Tukey’s post hoc test for multiple comparisons. The significance level was set at α < 0.05. Results: No statistically significant difference in irradiance was found between D1 (control) and D2. However, significantly lower values were recorded with A2, while D3, D4, and A1 produced significantly higher irradiance values compared to the control (p < 0.05). Conclusion: Irradiance measurements can vary significantly depending on the radiometer model used. Clinicians should be aware of this variability and are encouraged to regularly check the irradiance of the light-curing units used in daily practice, ensure their proper maintenance, and implement periodic monitoring to maintain effective clinical performance.

Authors

Institutions

Publication Details

Journal
Optics
Published
2025-06-30
DOI
https://doi.org/10.3390/opt6030028
Primary Topic
Impact of Light on Environment and Health
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Impact of Portable Radiometers on Irradiance Measurements of LED Photocuring Units

Carlos Enrique Cuevas‐Suárez, Alejandro Rivera-Gonzaga, Guillermo Grazioli, Rim Bourgi et al.
Optics
Impact of Light on Environment and Health
article

Impact of Portable Radiometers on Irradiance Measurements of LED Photocuring Units

Carlos Enrique Cuevas‐Suárez, Alejandro Rivera-Gonzaga, Guillermo Grazioli, Rim Bourgi, Andrés García, Matías Mederos, Elisa de León Cáceres
article en

Abstract

Purpose: The aim of this in vitro study was to evaluate the influence of different models of commercially available portable dental radiometers on the measurement of light irradiance emitted by light-emitting diode (LED) photocuring units. Materials and Methods: Eight LED photocuring units, all emitting light in a single-wavelength spectrum, were tested. Light irradiance (mW/cm2) was measured using six portable dental radiometers: four digital models (D1–D4) and two analog models (A1, A2). Digital model D1 was used as the reference (control). All measurements were conducted under standardized conditions, and each LED–radiometer combination was tested in triplicate. Data were analyzed using Sigma Plot 12.0 (Palo Alto, CA, USA) to verify the assumptions of normality and homogeneity of variances. A one-way analysis of variance (ANOVA) was used to assess the effect of the radiometer model on irradiance values, followed by Tukey’s post hoc test for multiple comparisons. The significance level was set at α < 0.05. Results: No statistically significant difference in irradiance was found between D1 (control) and D2. However, significantly lower values were recorded with A2, while D3, D4, and A1 produced significantly higher irradiance values compared to the control (p < 0.05). Conclusion: Irradiance measurements can vary significantly depending on the radiometer model used. Clinicians should be aware of this variability and are encouraged to regularly check the irradiance of the light-curing units used in daily practice, ensure their proper maintenance, and implement periodic monitoring to maintain effective clinical performance.

OpticsVol. 6(3)
Universidad Autónoma del Estado de Hidalgo (MX), Inserm (FR), Universidad de la República de Uruguay (UY), Saint Joseph University (LB), Beirut Arab University (LB), Université de Strasbourg (FR), Universidad de Montevideo (UY)
Openalex Percentile: Top 85%
Impact of Light on Environment and Health
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.