Transformer-Based Detection and Clinical Evaluation System for Torsional Nystagmus

<b>Motivation:</b> Benign paroxysmal positional vertigo (BPPV) is characterized by torsional nystagmus induced by changes in head position, where accurate quantitative assessment of subtle torsional eye movements is essential for precise diagnosis. Conventional videonystagmography (VNG) techniques face challenges in accurately capturing the rotational components of pupil movements, and existing automated methods typically exhibit limited performance in identifying torsional nystagmus. <b>Methodology:</b> The objective of this study was to develop an automated system capable of accurately and quantitatively detecting torsional nystagmus. We introduce the Torsion Transformer model, designed to directly estimate torsion angles from iris images. This model employs a self-supervised learning framework comprising two main components: a Decoder module, which learns rotational transformations from image data, and a Finder module, which subsequently estimates the torsion angle. The resulting torsion angle data, represented as time-series, are then analyzed using a 1-dimensional convolutional neural network (1D-CNN) classifier to detect the presence of nystagmus. The performance of the proposed method was evaluated using video recordings from 127 patients diagnosed with BPPV. <b>Findings:</b> Our Torsion Transformer model demonstrated robust performance, achieving a sensitivity of 89.99%, specificity of 86.36%, an F1-score of 88.82%, and an area under the receiver operating characteristic curve (AUROC) of 87.93%. These results indicate that the proposed model effectively quantifies torsional nystagmus, with performance levels comparable to established methods for detecting horizontal and vertical nystagmus. Thus, the Torsion Transformer shows considerable promise as a clinical decision support tool in the diagnosis of BPPV. <b>Key Findings:</b> Technical performance improvement in torsional nystagmus detection; System to support clinical decision-making for healthcare professionals.

Authors

Institutions

Publication Details

Journal
Sensors
Published
2025-06-28
DOI
https://doi.org/10.3390/s25134039
Primary Topic
Vestibular and auditory disorders
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Transformer-Based Detection and Clinical Evaluation System for Torsional Nystagmus

Jong‐Yeup Kim, Yong‐Suk Kim, Jong Bin Lee, Y. CHO et al.
Sensors
Vestibular and auditory disorders
article

Transformer-Based Detection and Clinical Evaluation System for Torsional Nystagmus

Jong‐Yeup Kim, Yong‐Suk Kim, Jong Bin Lee, Y. CHO, Hantai Kim, Ju Han
article en

Abstract

<b>Motivation:</b> Benign paroxysmal positional vertigo (BPPV) is characterized by torsional nystagmus induced by changes in head position, where accurate quantitative assessment of subtle torsional eye movements is essential for precise diagnosis. Conventional videonystagmography (VNG) techniques face challenges in accurately capturing the rotational components of pupil movements, and existing automated methods typically exhibit limited performance in identifying torsional nystagmus. <b>Methodology:</b> The objective of this study was to develop an automated system capable of accurately and quantitatively detecting torsional nystagmus. We introduce the Torsion Transformer model, designed to directly estimate torsion angles from iris images. This model employs a self-supervised learning framework comprising two main components: a Decoder module, which learns rotational transformations from image data, and a Finder module, which subsequently estimates the torsion angle. The resulting torsion angle data, represented as time-series, are then analyzed using a 1-dimensional convolutional neural network (1D-CNN) classifier to detect the presence of nystagmus. The performance of the proposed method was evaluated using video recordings from 127 patients diagnosed with BPPV. <b>Findings:</b> Our Torsion Transformer model demonstrated robust performance, achieving a sensitivity of 89.99%, specificity of 86.36%, an F1-score of 88.82%, and an area under the receiver operating characteristic curve (AUROC) of 87.93%. These results indicate that the proposed model effectively quantifies torsional nystagmus, with performance levels comparable to established methods for detecting horizontal and vertical nystagmus. Thus, the Torsion Transformer shows considerable promise as a clinical decision support tool in the diagnosis of BPPV. <b>Key Findings:</b> Technical performance improvement in torsional nystagmus detection; System to support clinical decision-making for healthcare professionals.

SensorsVol. 25(13)
Konyang University (KR), Konyang University Hospital (KR)
Korea Health Industry Development Institute
Peace, Justice and strong institutions
Openalex Percentile: Top 80%
Vestibular and auditory disorders
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.