Technical Validation of a Training Workstation for Magnet-Based Ultrasound Guidance of Fine-Needle Punctures

It has been demonstrated that needle guidance systems can enhance the precision and safety of ultrasound-guided punctures in human medicine. Systems that permit the utilization of commercially available standard needles, instead of those that necessitate the acquisition of costly, proprietary needles, are of particular interest. The objective of this phantom study is to evaluate the reliability and accuracy of magnet-based ultrasound needle guidance systems, which superimpose the position of the needle tip and a predictive trajectory line on the live ultrasound image. We conducted fine-needle aspiration cytology of thyroid nodules. The needles utilized in these procedures are of a slender gauge (21-27G), with lengths ranging from 40 to 80 mm. A dedicated training workstation with integrated software-based analyses of the movement of the needle tip was utilized in 240 standardized phantom punctures (angle: 45°; target depth: 20 mm). No system failures occurred, and the target achieved its aim in all cases. The analysis of the software revealed stable procedural parameters with minor relative deviations from the predefined reference values regarding the distance of needle tip movement (-4.2% to +6.7%), needle tilt (-6.4% to +9.6%), and penetration depth (-7.5% to +4.5%). These deviations appeared to increase with the use of thin needles and, to a lesser extent, long needles. They are attributed to the slight bending of the needle inside the (phantom) tissue. The training workstation we employed is thus suitable for use in educational settings. Nevertheless, in intricate clinical puncture scenarios-for instance, in the case of unfavorable localized small lesions near critical anatomical structures, particularly those involving thin needles-caution is advised, and the system should not be relied upon exclusively.

Authors

Institutions

Publication Details

Journal
Sensors
Published
2025-06-30
DOI
https://doi.org/10.3390/s25134102
Primary Topic
Radiology practices and education
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Technical Validation of a Training Workstation for Magnet-Based Ultrasound Guidance of Fine-Needle Punctures

Steffen Schrott, Martin G. Freesmeyer, Falk Gühne, Laura Schreiber et al.
Sensors
Radiology practices and education
article

Technical Validation of a Training Workstation for Magnet-Based Ultrasound Guidance of Fine-Needle Punctures

Steffen Schrott, Martin G. Freesmeyer, Falk Gühne, Laura Schreiber, Philipp Seifert, Christian Kühnel, Richard L. Popp
article en

Abstract

It has been demonstrated that needle guidance systems can enhance the precision and safety of ultrasound-guided punctures in human medicine. Systems that permit the utilization of commercially available standard needles, instead of those that necessitate the acquisition of costly, proprietary needles, are of particular interest. The objective of this phantom study is to evaluate the reliability and accuracy of magnet-based ultrasound needle guidance systems, which superimpose the position of the needle tip and a predictive trajectory line on the live ultrasound image. We conducted fine-needle aspiration cytology of thyroid nodules. The needles utilized in these procedures are of a slender gauge (21-27G), with lengths ranging from 40 to 80 mm. A dedicated training workstation with integrated software-based analyses of the movement of the needle tip was utilized in 240 standardized phantom punctures (angle: 45°; target depth: 20 mm). No system failures occurred, and the target achieved its aim in all cases. The analysis of the software revealed stable procedural parameters with minor relative deviations from the predefined reference values regarding the distance of needle tip movement (-4.2% to +6.7%), needle tilt (-6.4% to +9.6%), and penetration depth (-7.5% to +4.5%). These deviations appeared to increase with the use of thin needles and, to a lesser extent, long needles. They are attributed to the slight bending of the needle inside the (phantom) tissue. The training workstation we employed is thus suitable for use in educational settings. Nevertheless, in intricate clinical puncture scenarios-for instance, in the case of unfavorable localized small lesions near critical anatomical structures, particularly those involving thin needles-caution is advised, and the system should not be relied upon exclusively.

SensorsVol. 25(13)
Jena University Hospital (DE)
Deutsche Forschungsgemeinschaft
Openalex Percentile: Top 77%
Radiology practices and education
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.