Design and <i>in-silico</i> evaluation of a novel beamline for precision small animal pencil beam scanning delivery at clinical proton therapy facilities

<i>Objectives.</i>Preclinical small animal proton beam irradiation systems are increasingly in demand. However, the absence of dedicated systems comparable in precision to those in clinical settings presents a considerable hurdle to investigations in this field. To address this need, the SIRMIO project has developed a novel compact beam transport system configured to degrade and focus clinical proton beams. The beamline, about 1 m long, housed in an environment to minimize scatter, includes degraders, collimators, and a permanent magnet quadrupole triplet to focus protons degraded from clinical energies. It is tailored to transport focused proton beams within the energy range of 20 to 50 MeV, ideal for small animal preclinical studies. The flexibility of this beamline design allows achieving beam-spot sizes of 1 mm sigma at the isocenter for all focused energies, with the particle fluence and spot sizes being variable through dynamic adjustment of the collimator and magnetic lattice. 3D scanning of the target volume is possible due to lateral beam scanning integrated into this design, without the use of additional scanning dipole magnets.<i>Approach.</i>The beamline was optimized using an accelerator beam optics code, followed by a Monte Carlo (MC) model to account for beam-matter interactions. Using an experimentally validated clinical proton beam phase space as input, degraded beams are transported through the MC model. Outcomes are assessed for beam characteristics and dosimetric properties.<i>Main results.</i>Beams transported by our proposed beamline design are shown to result in dosimetric properties suitable for preclinical studies, while also emulating realistic clinically relevant beam delivery scenarios like pencil beam scanning. Compared to a similar-sized collimator-only beamline, this design enhances transmission and reduces secondary dose at the target due to absence of scattering elements nearby.<i>Significance.</i>The portable SIRMIO beamline offers a flexible, precise alternative to passive methods for adapting clinical proton beams for small animal irradiation, enhancing preclinical research with clinically relevant beam delivery and enabling experiments in conditions more closely matching clinical practice.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2025-06-30
DOI
https://doi.org/10.1088/1361-6560/adea08
Citations
1
Primary Topic
Radiation Therapy and Dosimetry
Type
article
Field-Weighted Citation Impact
1.52

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Design and in-silico evaluation of a novel beamline for precision small animal pencil beam scanning delivery at clinical proton therapy facilities

Katia Parodi, Neeraj Kurichiyanil, Marco Pinto
1 citations
Physics in Medicine and Biology
Radiation Therapy and Dosimetry
1.52
article

Design and in-silico evaluation of a novel beamline for precision small animal pencil beam scanning delivery at clinical proton therapy facilities

Katia Parodi, Neeraj Kurichiyanil, Marco Pinto
article en
1 citations

Abstract

<i>Objectives.</i>Preclinical small animal proton beam irradiation systems are increasingly in demand. However, the absence of dedicated systems comparable in precision to those in clinical settings presents a considerable hurdle to investigations in this field. To address this need, the SIRMIO project has developed a novel compact beam transport system configured to degrade and focus clinical proton beams. The beamline, about 1 m long, housed in an environment to minimize scatter, includes degraders, collimators, and a permanent magnet quadrupole triplet to focus protons degraded from clinical energies. It is tailored to transport focused proton beams within the energy range of 20 to 50 MeV, ideal for small animal preclinical studies. The flexibility of this beamline design allows achieving beam-spot sizes of 1 mm sigma at the isocenter for all focused energies, with the particle fluence and spot sizes being variable through dynamic adjustment of the collimator and magnetic lattice. 3D scanning of the target volume is possible due to lateral beam scanning integrated into this design, without the use of additional scanning dipole magnets.<i>Approach.</i>The beamline was optimized using an accelerator beam optics code, followed by a Monte Carlo (MC) model to account for beam-matter interactions. Using an experimentally validated clinical proton beam phase space as input, degraded beams are transported through the MC model. Outcomes are assessed for beam characteristics and dosimetric properties.<i>Main results.</i>Beams transported by our proposed beamline design are shown to result in dosimetric properties suitable for preclinical studies, while also emulating realistic clinically relevant beam delivery scenarios like pencil beam scanning. Compared to a similar-sized collimator-only beamline, this design enhances transmission and reduces secondary dose at the target due to absence of scattering elements nearby.<i>Significance.</i>The portable SIRMIO beamline offers a flexible, precise alternative to passive methods for adapting clinical proton beams for small animal irradiation, enhancing preclinical research with clinically relevant beam delivery and enabling experiments in conditions more closely matching clinical practice.

Physics in Medicine and BiologyVol. 70(14)
Ludwig-Maximilians-Universität München (DE)
H2020 European Research Council
Affordable and clean energy
Openalex Percentile: Top 17%
Radiation Therapy and Dosimetry
1.52
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.