METHODOLOGY FOR TESTING MEDICAL ROBOT AXES USING MULTIFACTORIAL TESTS WITH AN INERTIAL TEST BENCH

This work aims to develop a methodology for multi-factor testing of medical robot axes using an inertia test bench, which consists of a mechatronic module under test connected to a load mass driven by a motor through an electromagnetic clutch, along with a force-torque sensor and two optical position sensors. During the conducted experiments, the authors determine the characteristics of nominal braking parameters, as well as relationships between brake process parameters and given velocity and magnitude of external moment of inertia; collect data from the sensors and process it by an STM32 board, then record it in a text file format (.txt). Based on these findings, further adjustments can be made to the design parameters of the mechatronic axis module of a medical robot to modify its functional dependence and improve device performance.

Authors

Institutions

Publication Details

Journal
Automation and modeling in design and management of
Published
2025-06-30
DOI
https://doi.org/10.30987/2658-6436-2025-2-26-34
Primary Topic
Prosthetics and Rehabilitation Robotics
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

METHODOLOGY FOR TESTING MEDICAL ROBOT AXES USING MULTIFACTORIAL TESTS WITH AN INERTIAL TEST BENCH

Ilya Frolov, Denis Tarasov, Yuriy V. Poduraev, Andrey Vorotnikov
Automation and modeling in design and management of
Prosthetics and Rehabilitation Robotics
article

METHODOLOGY FOR TESTING MEDICAL ROBOT AXES USING MULTIFACTORIAL TESTS WITH AN INERTIAL TEST BENCH

Ilya Frolov, Denis Tarasov, Yuriy V. Poduraev, Andrey Vorotnikov
article en

Abstract

This work aims to develop a methodology for multi-factor testing of medical robot axes using an inertia test bench, which consists of a mechatronic module under test connected to a load mass driven by a motor through an electromagnetic clutch, along with a force-torque sensor and two optical position sensors. During the conducted experiments, the authors determine the characteristics of nominal braking parameters, as well as relationships between brake process parameters and given velocity and magnitude of external moment of inertia; collect data from the sensors and process it by an STM32 board, then record it in a text file format (.txt). Based on these findings, further adjustments can be made to the design parameters of the mechatronic axis module of a medical robot to modify its functional dependence and improve device performance.

Automation and modeling in design and management of
Lomonosov Moscow State University (RU), Moscow State Technological University (RU), Ministry of Health of the Russian Federation (RU), University of Technology (RU), Moscow State University (TJ)
Openalex Percentile: Top 86%
Prosthetics and Rehabilitation Robotics
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

METHODOLOGY FOR TESTING MEDICAL ROBOT AXES USING MULTIFACTORIAL TESTS WITH AN INERTIAL TEST BENCH — Ilya Frolov, Denis Tarasov, et al. · Automation and modeling in design and management of (2025) | TGRS Research Map | TGRS