Feasibility study of a multi‐lesion cyberknife radiotherapy plan verification method using a 2D array with pre‐set roll angles
SRSmapcheck with the pre-set roll angle method can verify intracranial multi-lesion CyberKnife plans by measuring multiple targets in a single validation and comparing the 1 mm/1% and 1 mm/2% γ analysis criteria.
Authors
- Jinhu Chen (ORCID: https://orcid.org/0000-0002-4923-5076)
- Tengxiang Li (ORCID: https://orcid.org/0000-0003-4320-0363)
- Quan Tang (ORCID: https://orcid.org/0000-0002-2365-2840)
- Rui He (ORCID: https://orcid.org/0000-0002-8314-9155)
- Yong Yin (ORCID: https://orcid.org/0000-0003-2321-9200)
- Qingtao Qiu (ORCID: https://orcid.org/0000-0002-8355-0125)
Institutions
- Shandong Tumor Hospital (CN)
- Shandong First Medical University (CN)
- University of South China (CN)
Publication Details
- Journal
- Precision Radiation Oncology
- Published
- 2025-06-28
- DOI
- https://doi.org/10.1002/pro6.70022
- Primary Topic
- Advanced Radiotherapy Techniques
- Type
- article
- Field-Weighted Citation Impact
- 0.00
Funders
- National Natural Science Foundation of China