Туяа эмчилгээний шугаман хурдасгуурын гүнд шингэсэн тунгийн харьцааг PHITS кодоор тооцоолсон үр дүн
In radiation therapy, the Percentage Depth Dose (PDD) is a key parameter for evaluating dose distribution, playing a critical role in treatment planning and quality assurance. In this study, the PDD of 6 MV photon beams produced by a Varian TrueBeam linear accelerator was calculated using the PHITS code. The simulated results were then compared with measured values and standard value. The findings indicate that the PHITS-calculated PDD values differed by ±2% within the 150–200 mm depth range, which falls within the acceptable tolerance limits recommended by the AAPM TG-40 report.
Authors
- Ganzul Tsengel
Publication Details
- Journal
- Физик сэтгүүл
- Published
- 2025-01-01
- DOI
- https://doi.org/10.22353/physics.v38i621.10096
- Primary Topic
- Advanced Radiotherapy Techniques
- Type
- article
- Field-Weighted Citation Impact
- 0.00