Water calorimetry-based determination of absorbed dose to water and beam quality correction factors for ten ionization chambers in scanned proton beams

<i>Objective.</i>To experimentally determine beam quality correction factors (<i>k<sub>Q</sub></i>) for six cylindrical and four parallel-plate ionization chambers (ICs) in both spread-out Bragg peak (SOBP) and single-layer proton beam configurations.<i>Approach.</i>Water calorimetry was implemented to establish absorbed dose to water (<i>D</i><sub>w</sub>) at 10 g cm<sup>-2</sup>depth for SOBP and single-layer proton beams. IC measurements were performed under identical geometrical conditions as calorimetric measurements, with the exception of water temperature control in the phantom. Systematic evaluation of ion recombination and polarity effects was achieved through sequential measurements at operational voltages of -400 V, -300 V, -200 V, -150 V, and +400 V in scanned proton beams. All chambers were calibrated against the<sup>60</sup>Co<i>D</i><sub>w</sub>standard at the National Institute of Metrology to obtain<i>D</i><sub>w</sub>calibration coefficients (<i>N</i><sub><i>D</i>,w</sub>). The<i>k<sub>Q</sub></i>values were determined through integrated water calorimetry and IC measurements.<i>Main results.</i>The relative standard uncertainties in<i>D</i><sub>w</sub>determination were quantified as 0.43% for SOBP beams and 0.51% for single-layer beams. The experimentally determined<i>k<sub>Q</sub></i>values demonstrated agreement with both TRS-398 reference data and published literature within declared measurement uncertainties.<i>Significance.</i>This work presents the first comprehensive determination of<i>k<sub>Q</sub></i>factors for SOBP and single-layer proton beams using water calorimetry. The obtained<i>k<sub>Q</sub></i>values with reduced uncertainties (0.56%-0.63%) establish metrological traceability for ten clinical IC models, directly enhancing the accuracy of scanned proton beam dosimetry in radiotherapy practice.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2025-06-27
DOI
https://doi.org/10.1088/1361-6560/ade8cf
Citations
1
Primary Topic
Radiation Therapy and Dosimetry
Type
article
Field-Weighted Citation Impact
1.52

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Water calorimetry-based determination of absorbed dose to water and beam quality correction factors for ten ionization chambers in scanned proton beams

Han Wu, Jiale Han, Jinhu Chen, Jintao Yao et al.
1 citations
Physics in Medicine and Biology
Radiation Therapy and Dosimetry
1.52
article

Water calorimetry-based determination of absorbed dose to water and beam quality correction factors for ten ionization chambers in scanned proton beams

Han Wu, Jiale Han, Jinhu Chen, Jintao Yao, Sunjun Jin, Yaping Qi, Kun Wang, Zhipeng Wang, Jian Zhu, Ji Huang, Gang Lü, Chengqiang Li, Yungang Wang
article en
1 citations

Abstract

<i>Objective.</i>To experimentally determine beam quality correction factors (<i>k<sub>Q</sub></i>) for six cylindrical and four parallel-plate ionization chambers (ICs) in both spread-out Bragg peak (SOBP) and single-layer proton beam configurations.<i>Approach.</i>Water calorimetry was implemented to establish absorbed dose to water (<i>D</i><sub>w</sub>) at 10 g cm<sup>-2</sup>depth for SOBP and single-layer proton beams. IC measurements were performed under identical geometrical conditions as calorimetric measurements, with the exception of water temperature control in the phantom. Systematic evaluation of ion recombination and polarity effects was achieved through sequential measurements at operational voltages of -400 V, -300 V, -200 V, -150 V, and +400 V in scanned proton beams. All chambers were calibrated against the<sup>60</sup>Co<i>D</i><sub>w</sub>standard at the National Institute of Metrology to obtain<i>D</i><sub>w</sub>calibration coefficients (<i>N</i><sub><i>D</i>,w</sub>). The<i>k<sub>Q</sub></i>values were determined through integrated water calorimetry and IC measurements.<i>Main results.</i>The relative standard uncertainties in<i>D</i><sub>w</sub>determination were quantified as 0.43% for SOBP beams and 0.51% for single-layer beams. The experimentally determined<i>k<sub>Q</sub></i>values demonstrated agreement with both TRS-398 reference data and published literature within declared measurement uncertainties.<i>Significance.</i>This work presents the first comprehensive determination of<i>k<sub>Q</sub></i>factors for SOBP and single-layer proton beams using water calorimetry. The obtained<i>k<sub>Q</sub></i>values with reduced uncertainties (0.56%-0.63%) establish metrological traceability for ten clinical IC models, directly enhancing the accuracy of scanned proton beam dosimetry in radiotherapy practice.

Physics in Medicine and BiologyVol. 70(14)
Anhui University of Science and Technology (CN), Shandong Tumor Hospital (CN), National Institute of Metrology (CN), Shandong First Medical University (CN)
Natural Science Foundation of Shandong Province
Clean water and sanitation
Openalex Percentile: Top 17%
Radiation Therapy and Dosimetry
1.52
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.