An alternative real-time fluorescence reverse transcription loop-mediated isothermal amplification assay for the rapid detection of SARS-CoV-2

COVID-19, caused by SARS-CoV-2 virus infection, remains a public health concern in many countries. Reverse transcription loop-mediated isothermal amplification (RT-LAMP) is a rapid and cost-effective alternative test for COVID-19 diagnosis. In this study, we developed and evaluated a real-time RT-LAMP (rRT-LAMP) assay coupled with a melting curve analysis to detect SARS-CoV-2. The reaction was carried out in a real-time thermal cycler at 63 °C for 45 min to amplify the region of SARS-CoV-2 orf8; real-time monitoring of amplification was performed by fluorescence detection. The performance was assessed by comparing it to a real-time reverse transcription-polymerase reaction (rRT-PCR) as a reference. The rRT-LAMP could detect as few as 15 copies of SARS-CoV-2 RNA per reaction. Positive results appeared within 30 min, while the melting-temperature analysis could verify the amplification specificity. No positive results from non-SARS-CoV-2 templates and no mis-amplification were observed. The comparative analysis using 262 RNA extracted from nasopharyngeal swab samples revealed the overall accuracy, sensitivity, specificity, positive predictive value (PPV), and negative predictive values (NPV) of the rRT-LAMP at 88.55% (95% CI: 77.52-100%), 84.13% (95% CI: 71.56-98.27%), 100% (95% CI: 78.38-100%), 100% (95% CI: 85.06-100%), and 70.87% (95% CI: 55.55-89.11%), respectively. The greatest sensitivity was as high as 98-100% for specimens with threshold rRT-PCR cycle (Ct) values of less than 30 cycles. Overall, this rRT-LAMP showed good performance for the rapid detection of SARS-CoV-2. It is proposed as a potential method for real-time amplification detection, offering increased laboratory capacity for SARS-CoV-2 testing in a cost-effective and timely manner.

Authors

Institutions

Publication Details

Journal
Revista do Instituto de Medicina Tropical de São Paulo
Published
2025-01-01
DOI
https://doi.org/10.1590/s1678-9946202567037
Citations
1
Primary Topic
Biosensors and Analytical Detection
Type
article
Field-Weighted Citation Impact
0.67

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

An alternative real-time fluorescence reverse transcription loop-mediated isothermal amplification assay for the rapid detection of SARS-CoV-2

Benjawan Phetsuksiri, Janisara Rudeeaneksin, Nuttagarn Chuenchom, Wiphat Klayut et al.
1 citations
Revista do Instituto de Medicina Tropical de São Paulo
Biosensors and Analytical Detection
0.67
article

An alternative real-time fluorescence reverse transcription loop-mediated isothermal amplification assay for the rapid detection of SARS-CoV-2

Benjawan Phetsuksiri, Janisara Rudeeaneksin, Nuttagarn Chuenchom, Wiphat Klayut, Thanee Wongchai, Ballang Uppapong
article en
1 citations

Abstract

COVID-19, caused by SARS-CoV-2 virus infection, remains a public health concern in many countries. Reverse transcription loop-mediated isothermal amplification (RT-LAMP) is a rapid and cost-effective alternative test for COVID-19 diagnosis. In this study, we developed and evaluated a real-time RT-LAMP (rRT-LAMP) assay coupled with a melting curve analysis to detect SARS-CoV-2. The reaction was carried out in a real-time thermal cycler at 63 °C for 45 min to amplify the region of SARS-CoV-2 orf8; real-time monitoring of amplification was performed by fluorescence detection. The performance was assessed by comparing it to a real-time reverse transcription-polymerase reaction (rRT-PCR) as a reference. The rRT-LAMP could detect as few as 15 copies of SARS-CoV-2 RNA per reaction. Positive results appeared within 30 min, while the melting-temperature analysis could verify the amplification specificity. No positive results from non-SARS-CoV-2 templates and no mis-amplification were observed. The comparative analysis using 262 RNA extracted from nasopharyngeal swab samples revealed the overall accuracy, sensitivity, specificity, positive predictive value (PPV), and negative predictive values (NPV) of the rRT-LAMP at 88.55% (95% CI: 77.52-100%), 84.13% (95% CI: 71.56-98.27%), 100% (95% CI: 78.38-100%), 100% (95% CI: 85.06-100%), and 70.87% (95% CI: 55.55-89.11%), respectively. The greatest sensitivity was as high as 98-100% for specimens with threshold rRT-PCR cycle (Ct) values of less than 30 cycles. Overall, this rRT-LAMP showed good performance for the rapid detection of SARS-CoV-2. It is proposed as a potential method for real-time amplification detection, offering increased laboratory capacity for SARS-CoV-2 testing in a cost-effective and timely manner.

Revista do Instituto de Medicina Tropical de São PauloVol. 67
Ministry of Public Health (TH), Maesot General Hospital (TH), National Institute of Health of Thailand (TH)
National Research Council of Thailand
Good health and well-being, Partnerships for the goals
Openalex Percentile: Top 30%
Biosensors and Analytical Detection
0.67
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.