Development of a headgear-based eye protection device for physicians performing fluoroscopy-guided bronchoscopy

Fluoroscopy-guided bronchoscopy is an essential tool for diagnosing and treating lung diseases, particularly lung cancer. However, prolonged fluoroscopic exposure raises concerns regarding radiation-induced lens injuries in physicians, such as radiation cataracts. In response to the International Commission on Radiological Protection lowering the occupational lens dose limit to an average of 20 mSv/year over 5 years, there is an increasing need for effective lens protection during such procedures. This study has aimed to develop a novel lens protection device specifically designed for bronchoscopy physicians and evaluate its protective performance through a phantom study. The device consisted of a 0.175 mm lead (Pb) sheet positioned on the left side of the physician's head, secured with headgear to improve stability and comfort during prolonged use. A phantom study was conducted using a trunk phantom to simulate a patient and a head phantom to simulate a physician. The scattered radiation doses were measured at 15 locations on the phantom head using a radiophotoluminescence glass dosimeter, both with and without a protective device. The device demonstrated a protective effect of more than 80% for the left eye across all tested angles, whereas the right eye showed protection ranging from approximately 40% to 75% depending on the angle. This novel lens protection device has the potential to significantly reduce scattered radiation to the left eye while minimizing vision obstruction and discomfort. This offers a practical solution for radiation protection during bronchoscopy and may be applicable to other interventional procedures requiring fluoroscopic guidance.

Authors

Institutions

Publication Details

Journal
Journal of Radiation Research
Published
2025-06-18
DOI
https://doi.org/10.1093/jrr/rraf031
Citations
3
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
4.08

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Development of a headgear-based eye protection device for physicians performing fluoroscopy-guided bronchoscopy

Yohei Inaba, Koichi Chida, Yoshihiro Haga, Masahiro Sota et al.
3 citations
Journal of Radiation Research
Radiation Dose and Imaging
4.08
article

Development of a headgear-based eye protection device for physicians performing fluoroscopy-guided bronchoscopy

Yohei Inaba, Koichi Chida, Yoshihiro Haga, Masahiro Sota, Toshiki Kato, Masaki Fujisawa, Saki Takahira, Masatoshi Suzuki, Yuji Kaga, Mitsuya Abe
article en
3 citations

Abstract

Fluoroscopy-guided bronchoscopy is an essential tool for diagnosing and treating lung diseases, particularly lung cancer. However, prolonged fluoroscopic exposure raises concerns regarding radiation-induced lens injuries in physicians, such as radiation cataracts. In response to the International Commission on Radiological Protection lowering the occupational lens dose limit to an average of 20 mSv/year over 5 years, there is an increasing need for effective lens protection during such procedures. This study has aimed to develop a novel lens protection device specifically designed for bronchoscopy physicians and evaluate its protective performance through a phantom study. The device consisted of a 0.175 mm lead (Pb) sheet positioned on the left side of the physician's head, secured with headgear to improve stability and comfort during prolonged use. A phantom study was conducted using a trunk phantom to simulate a patient and a head phantom to simulate a physician. The scattered radiation doses were measured at 15 locations on the phantom head using a radiophotoluminescence glass dosimeter, both with and without a protective device. The device demonstrated a protective effect of more than 80% for the left eye across all tested angles, whereas the right eye showed protection ranging from approximately 40% to 75% depending on the angle. This novel lens protection device has the potential to significantly reduce scattered radiation to the left eye while minimizing vision obstruction and discomfort. This offers a practical solution for radiation protection during bronchoscopy and may be applicable to other interventional procedures requiring fluoroscopic guidance.

Journal of Radiation ResearchVol. 66(4)
Tohoku University (JP), Sendai Kousei Hospital (JP)
Japan Society for the Promotion of Science
Good health and well-being
Openalex Percentile: Top 6%
Radiation Dose and Imaging
4.08
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.