<sup>241</sup> Am Radioisotope-Excited Spectrometer for K-Line X-ray Fluorescence (XRF) Analysis of Rare Earth Elements

We describe a radioisotope-excited energy dispersive X-ray fluorescence spectrometric facility for the elemental analysis of minerals containing the rare earth elements (REEs) by the use of the K-line X-rays. Two <sup>241</sup>Am sources are used in purpose-designed holders that fit a high-resolution solid-state germanium detector. The system is capable of exciting all the light actinides up to element 69, Tm. The analysis through the well-separated K-lines allows an easy identification of lanthanide elements with advantages over traditional L-line detection particularly in mineral samples with high Fe concentration, enabling the recognition of most lanthanide elements at values close to 10 ppm. We report the achieved elemental sensitivity obtained with a set of pure element standards and spectra of mineral concentrates derived from a REEcontaining Venezuelan Laterite.

Authors

Institutions

Publication Details

Journal
Applied Spectroscopy
Published
2025-06-27
DOI
https://doi.org/10.1177/00037028251348483
Primary Topic
X-ray Spectroscopy and Fluorescence Analysis
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

241 Am Radioisotope-Excited Spectrometer for K-Line X-ray Fluorescence (XRF) Analysis of Rare Earth Elements

Monserrat Gutierrez, Sanber Vizcaya, E. D. Greaves
Applied Spectroscopy
X-ray Spectroscopy and Fluorescence Analysis
article

241 Am Radioisotope-Excited Spectrometer for K-Line X-ray Fluorescence (XRF) Analysis of Rare Earth Elements

Monserrat Gutierrez, Sanber Vizcaya, E. D. Greaves
article en

Abstract

We describe a radioisotope-excited energy dispersive X-ray fluorescence spectrometric facility for the elemental analysis of minerals containing the rare earth elements (REEs) by the use of the K-line X-rays. Two <sup>241</sup>Am sources are used in purpose-designed holders that fit a high-resolution solid-state germanium detector. The system is capable of exciting all the light actinides up to element 69, Tm. The analysis through the well-separated K-lines allows an easy identification of lanthanide elements with advantages over traditional L-line detection particularly in mineral samples with high Fe concentration, enabling the recognition of most lanthanide elements at values close to 10 ppm. We report the achieved elemental sensitivity obtained with a set of pure element standards and spectra of mineral concentrates derived from a REEcontaining Venezuelan Laterite.

Applied SpectroscopyVol. 79(12)
Simón Bolívar University (VE), Federico Santa María Technical University (CL), Universidad Central de Venezuela (VE)
Fondo Nacional de Ciencia Tecnología e Innovación
Openalex Percentile: Top 78%
X-ray Spectroscopy and Fluorescence Analysis
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

<sup>241</sup> Am Radioisotope-Excited Spectrometer for K-Line X-ray Fluorescence (XRF) Analysis of Rare Earth Elements — Monserrat Gutierrez, Sanber Vizcaya, et al. · Applied Spectroscopy (2025) | TGRS Research Map | TGRS