A programmable charge-balanced current stimulation circuit for implantable medical devices

Active implantable medical devices (IMDs) require precise tissue stimulation to achieve effective therapeutic outcomes and ensure patient safety. This work introduces an 8-bit programmable current source, capable of delivering biphasic stimulation pulses from -25.5 mA to +25.5 mA in 0.1 mA steps, suitable for common IMD applications. The circuit supports pulses as short as a few microseconds, with measured settling times under 1 μs. The system includes a charge mismatch measurement circuit, and a 5-bit trimming mechanism to minimize the net charge delivered to the tissue. Fabricated in a 180 nm high-voltage CMOS technology, the circuit occupies less than 4 mm<sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sup>, including a RISC-V processor, and I/O pads. Measurement results from the fabricated circuit are provided to validate the design.

Authors

Publication Details

Published
2025-05-25
DOI
https://doi.org/10.1109/iscas56072.2025.11044212
Primary Topic
Neuroscience and Neural Engineering
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A programmable charge-balanced current stimulation circuit for implantable medical devices

Alfredo Arnaud, Leonardo Agis, Matías Miguez, Natalia J. Martinez et al.
Neuroscience and Neural Engineering
article

A programmable charge-balanced current stimulation circuit for implantable medical devices

Alfredo Arnaud, Leonardo Agis, Matías Miguez, Natalia J. Martinez, Augusto Sbarbaro-Benech
article en

Abstract

Active implantable medical devices (IMDs) require precise tissue stimulation to achieve effective therapeutic outcomes and ensure patient safety. This work introduces an 8-bit programmable current source, capable of delivering biphasic stimulation pulses from -25.5 mA to +25.5 mA in 0.1 mA steps, suitable for common IMD applications. The circuit supports pulses as short as a few microseconds, with measured settling times under 1 μs. The system includes a charge mismatch measurement circuit, and a 5-bit trimming mechanism to minimize the net charge delivered to the tissue. Fabricated in a 180 nm high-voltage CMOS technology, the circuit occupies less than 4 mm<sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sup>, including a RISC-V processor, and I/O pads. Measurement results from the fabricated circuit are provided to validate the design.

Openalex Percentile: Top 84%
Neuroscience and Neural Engineering
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.