A Failure Detection Technique for Au-Plated CMOS Microelectrode Arrays
A method for the detection of failure in gold-plated microelectrodes is presented. Three main failure mechanisms–layer detachment, biofouling and halide-mediated pas-sivation (HMP)–were induced on the microelectrodes, leading to electrode failure. Sequential admittance and sensitivity measurements were carried out on gold-plated microelectrodes on printed circuit boards (PCBs) and CMOS chips to develop and train the classifier. A two-step classifier is proposed that relies only on measuring the admittance magnitude (|Y|) at three specific frequencies within 0.6-800 Hz. The classifier achieves 88% accuracy on PCB microelectrodes and 92% accuracy on CMOS microelectrodes.
Authors
- Sara S. Ghoreishizadeh (ORCID: https://orcid.org/0000-0002-2398-0887)
- Minghao Li (ORCID: https://orcid.org/0000-0002-4655-1217)
- Xin Zhang (ORCID: https://orcid.org/0000-0002-6669-7037)
Institutions
- University College London (GB)
Publication Details
- Published
- 2025-05-25
- DOI
- https://doi.org/10.1109/iscas56072.2025.11043579
- Citations
- 1
- Primary Topic
- Integrated Circuits and Semiconductor Failure Analysis
- Type
- article
- Field-Weighted Citation Impact
- 0.73