Photon-counting micro-CT scanner for deep learning-enabled small animal perfusion imaging

<i>Objective.</i>In this work, we introduce a benchtop, turn-table photon-counting (PC) micro-computed tomography (CT) scanner and highlight its application for dynamic small animal perfusion imaging.<i>Approach.</i>Built on recently published hardware, the system now features a CdTe-based PC detector. We validated its static spectral PC micro-CT imaging using conventional phantoms and assessed dynamic performance with a custom flow-configurable dual-compartment perfusion phantom. The phantom was scanned under varied flow conditions during injections of a low molecular weight iodinated contrast agent.<i>In vivo</i>mouse studies with identical injection settings demonstrated potential applications. A pretrained denoising convolutional neural network (CNN) processed large multi-energy, temporal datasets (20 timepoints ×4 energies ×3 spatial dimensions), reconstructed via weighted filtered back projection. A separate CNN, trained on simulated data, performed gamma variate-based 2D perfusion mapping, evaluated qualitatively in phantom and<i>in vivo</i>tests.<i>Main Results.</i>Full five-dimensional reconstructions were denoised using a CNN in ∼3% of the time of iterative reconstruction, reducing noise in water at the highest energy threshold from 1206 HU to 86 HU. Decomposed iodine maps, which improved contrast to noise ratio from 16.4 (in the lowest energy CT images) to 29.4 (in the iodine maps), were used for perfusion analysis. The perfusion CNN outperformed pixelwise gamma variate fitting by ∼33%, with a test set error of 0.04 vs. 0.06 in blood flow index (BFI) maps, and quantified linear BFI changes in the phantom with a coefficient of determination of 0.98.<i>Significance.</i>This work underscores the PC micro-CT scanner's utility for high-throughput small animal perfusion imaging, leveraging spectral PC micro-CT and iodine decomposition. It provides a versatile platform for preclinical vascular research and advanced, time-resolved studies of disease models and therapeutic interventions.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2025-06-27
DOI
https://doi.org/10.1088/1361-6560/ade94b
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Photon-counting micro-CT scanner for deep learning-enabled small animal perfusion imaging

Cristian T. Badea, Darin P. Clark, Rohan Nadkarni, Alex J. Allphin
1 links
Physics in Medicine and Biology
Advanced X-ray and CT Imaging
article

Photon-counting micro-CT scanner for deep learning-enabled small animal perfusion imaging

Cristian T. Badea, Darin P. Clark, Rohan Nadkarni, Alex J. Allphin
article en

Abstract

<i>Objective.</i>In this work, we introduce a benchtop, turn-table photon-counting (PC) micro-computed tomography (CT) scanner and highlight its application for dynamic small animal perfusion imaging.<i>Approach.</i>Built on recently published hardware, the system now features a CdTe-based PC detector. We validated its static spectral PC micro-CT imaging using conventional phantoms and assessed dynamic performance with a custom flow-configurable dual-compartment perfusion phantom. The phantom was scanned under varied flow conditions during injections of a low molecular weight iodinated contrast agent.<i>In vivo</i>mouse studies with identical injection settings demonstrated potential applications. A pretrained denoising convolutional neural network (CNN) processed large multi-energy, temporal datasets (20 timepoints ×4 energies ×3 spatial dimensions), reconstructed via weighted filtered back projection. A separate CNN, trained on simulated data, performed gamma variate-based 2D perfusion mapping, evaluated qualitatively in phantom and<i>in vivo</i>tests.<i>Main Results.</i>Full five-dimensional reconstructions were denoised using a CNN in ∼3% of the time of iterative reconstruction, reducing noise in water at the highest energy threshold from 1206 HU to 86 HU. Decomposed iodine maps, which improved contrast to noise ratio from 16.4 (in the lowest energy CT images) to 29.4 (in the iodine maps), were used for perfusion analysis. The perfusion CNN outperformed pixelwise gamma variate fitting by ∼33%, with a test set error of 0.04 vs. 0.06 in blood flow index (BFI) maps, and quantified linear BFI changes in the phantom with a coefficient of determination of 0.98.<i>Significance.</i>This work underscores the PC micro-CT scanner's utility for high-throughput small animal perfusion imaging, leveraging spectral PC micro-CT and iodine decomposition. It provides a versatile platform for preclinical vascular research and advanced, time-resolved studies of disease models and therapeutic interventions.

Physics in Medicine and BiologyVol. 70(14)
Duke University (US), Duke Medical Center (US)
Openalex Percentile: Top 86%
Advanced X-ray and CT Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.