Determinants of photoplethysmography signal quality at the wrist

Wrist photoplethysmogram (PPG) signals are widely used for physiological monitoring in consumer devices. However, the PPG is highly susceptible to noise, which can reduce the accuracy of monitored parameters. The aim of this study was to identify factors which influence PPG signal quality. Data from the Aurora-BP dataset were used, consisting of reflectance wrist PPG signals measured from 1,142 subjects of varying ages and health statuses. Measurements were acquired in supine, sitting, and standing postures, and with the sensor held at different heights. Three signal quality metrics were calculated: the signal-to-noise ratio (SNR), the perfusion index (PI), and the template-matching correlation coefficient (TMCC). When comparing between postures with the sensor held at a natural height, quality was greatest in the supine position (SNR: 18.6 dB), followed by sitting with the arm resting in the lap (13.7 dB), and lowest whilst standing with the arm hanging alongside (9.0 dB) (p < 0.001). Signal quality increased as the arm was raised to heart height: whilst sitting, quality was lowest with the arm alongside the body (10.5 dB), and increased when the sensor was held in the lap (13.7 dB) and at heart height (15.5 dB) (p < 0.001). Similar trends were observed for the TMCC and PI. Findings were mixed for the influence of participant characteristics on signal quality. The SNR and TMCC, but not the PI, increased with age. The SNR either decreased or remained constant at darker skin tones when controlling for PPG DC amplitude, compared to constant or increased when allowing DC amplitude to vary. In conclusion, this study identified the impacts of posture and sensor height on signal quality, with highest qualities observed in the supine posture and with the sensor at heart height. It also highlights the importance of adjusting LED light intensity to maintain signal quality across skin tones.

Authors

Institutions

Publication Details

Journal
PLOS Digital Health
Published
2025-06-27
DOI
https://doi.org/10.1371/journal.pdig.0000585
Citations
9
Primary Topic
Non-Invasive Vital Sign Monitoring
Type
article
Field-Weighted Citation Impact
6.07

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Determinants of photoplethysmography signal quality at the wrist

Vaidotas Marozas, Elisa Mejía‐Mejía, Jonathan Mant, P. A. Kyriacou et al.
9 citations
PLOS Digital Health
Non-Invasive Vital Sign Monitoring
6.07
article

Determinants of photoplethysmography signal quality at the wrist

Vaidotas Marozas, Elisa Mejía‐Mejía, Jonathan Mant, P. A. Kyriacou, Peter Charlton
article en
9 citations

Abstract

Wrist photoplethysmogram (PPG) signals are widely used for physiological monitoring in consumer devices. However, the PPG is highly susceptible to noise, which can reduce the accuracy of monitored parameters. The aim of this study was to identify factors which influence PPG signal quality. Data from the Aurora-BP dataset were used, consisting of reflectance wrist PPG signals measured from 1,142 subjects of varying ages and health statuses. Measurements were acquired in supine, sitting, and standing postures, and with the sensor held at different heights. Three signal quality metrics were calculated: the signal-to-noise ratio (SNR), the perfusion index (PI), and the template-matching correlation coefficient (TMCC). When comparing between postures with the sensor held at a natural height, quality was greatest in the supine position (SNR: 18.6 dB), followed by sitting with the arm resting in the lap (13.7 dB), and lowest whilst standing with the arm hanging alongside (9.0 dB) (p < 0.001). Signal quality increased as the arm was raised to heart height: whilst sitting, quality was lowest with the arm alongside the body (10.5 dB), and increased when the sensor was held in the lap (13.7 dB) and at heart height (15.5 dB) (p < 0.001). Similar trends were observed for the TMCC and PI. Findings were mixed for the influence of participant characteristics on signal quality. The SNR and TMCC, but not the PI, increased with age. The SNR either decreased or remained constant at darker skin tones when controlling for PPG DC amplitude, compared to constant or increased when allowing DC amplitude to vary. In conclusion, this study identified the impacts of posture and sensor height on signal quality, with highest qualities observed in the supine posture and with the sensor at heart height. It also highlights the importance of adjusting LED light intensity to maintain signal quality across skin tones.

PLOS Digital HealthVol. 4(6)
Rush University Medical Center (US), Kaunas University of Technology (LT), City, University of London (GB), University of Cambridge (GB)
European Partnership on Metrology, British Heart Foundation, Innovate UK
Openalex Percentile: Top 3%
Non-Invasive Vital Sign Monitoring
6.07
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.