Performance evaluation of the semi-automated urine analyser DFI R-600S

Objective: The major considerations of the urine analysis performed in laboratories are to obtain a fast and accurate test result. Increasing number of instruments and methods have been developed for making it possible. The aim of this study is to compare the performance of the DFI R600S with that of the Roche Cobas U411. Methods: Five hundred thirty three freshly obtained out- and in-patient routine urine samples collected. We conducted an analysis using results from urine dipstick tests that measure ten parameters: glucose, protein, bilirubin, urobilinogen, pH, specific gravity, ketones, red blood cells, nitrite, and leukocytes. Both analyzers were utilized in the process, followed by a microscopic examination of the urinary sediment. Results: It was determined that the pH and specific gravity tests showed poor concordance, while the remaining tests exhibited moderate to very good concordance between the manual microscopic method and the individual devices. Conclusion: The semi-automated test strip analyzer DFI R600S offers low cost, easy to use and reliable first level screening method for urinalysis but it is important to be aware of conditions produce false-positive or false-negative results of the urine dipstick.

Authors

Institutions

Publication Details

Journal
Clinical and Experimental Health Sciences
Published
2025-06-27
DOI
https://doi.org/10.33808/clinexphealthsci.1583681
Primary Topic
Healthcare Technology and Patient Monitoring
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Performance evaluation of the semi-automated urine analyser DFI R-600S

Murat Can, Berrak Güven, Ferhan Adıbelli, Emrah Altunsoy
Clinical and Experimental Health Sciences
Healthcare Technology and Patient Monitoring
article

Performance evaluation of the semi-automated urine analyser DFI R-600S

Murat Can, Berrak Güven, Ferhan Adıbelli, Emrah Altunsoy
article en

Abstract

Objective: The major considerations of the urine analysis performed in laboratories are to obtain a fast and accurate test result. Increasing number of instruments and methods have been developed for making it possible. The aim of this study is to compare the performance of the DFI R600S with that of the Roche Cobas U411. Methods: Five hundred thirty three freshly obtained out- and in-patient routine urine samples collected. We conducted an analysis using results from urine dipstick tests that measure ten parameters: glucose, protein, bilirubin, urobilinogen, pH, specific gravity, ketones, red blood cells, nitrite, and leukocytes. Both analyzers were utilized in the process, followed by a microscopic examination of the urinary sediment. Results: It was determined that the pH and specific gravity tests showed poor concordance, while the remaining tests exhibited moderate to very good concordance between the manual microscopic method and the individual devices. Conclusion: The semi-automated test strip analyzer DFI R600S offers low cost, easy to use and reliable first level screening method for urinalysis but it is important to be aware of conditions produce false-positive or false-negative results of the urine dipstick.

Clinical and Experimental Health SciencesVol. 15(2)
Zonguldak Bülent Ecevit University (TR)
Openalex Percentile: Top 74%
Healthcare Technology and Patient Monitoring
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.