Correction of X-ray tube aging effects and quantitative estimation of elemental composition using the ITRAX XRF core scanner: A case study of Japan Sea sediments

The XRF core scanner, which facilitates non-destructive, high-speed, and high spatial resolution measurements, has recently been extensively used to measure element variability in sediments. Although identifying the precision of the measurement and concentration of the elements in the sample is essential to elucidating the variations recorded in the sediments, the element peak area count ratios are most commonly used to report the results of XRF core scanners in many previous studies, and element concentration is not discussed. This is mainly because quantitative calibration of element concentration necessitates numerous additional analyses of discrete samples. Because XRF core scanner measurements are often conducted before discrete sample analysis, if approximate element concentrations in the sediment can be estimated from XRF core scanner data just after measurement, sampling intervals should be determined for further analysis.In this study, we developed a new database on the aging of X-ray tubes, and propose a method for estimating element concentrations in the wet sediments from XRF core scanner measurement without any additional measurement. The ITRAX XRF core scanner at Kochi University, the first ITRAX installed in Japan in 2014, was used in this study. Mo X-ray tubes were used at 30 kV, 55 mA with XRF exposure time of 10 s or 32 s. First, the aging effect of the X-ray tube was monitored and its correction method was proposed. Subsequently, the precision of the ITRAX XRF measurements was evaluated and the relationship between the element peak area count by ITRAX and element concentration was analyzed. Finally, we propose a new method for applying these databases constructed in this study to any results measured by ITRAX XRF core scanner at Kochi University. Using this method, anyone can estimate approximate element composition of the sediment from their XRF core scanner data obtained at Kochi University without any additional measurement. This method even enables estimation of element concentrations of sediments just after the ITRAX measurement.

Authors

Institutions

Publication Details

Journal
GEOCHEMICAL JOURNAL
Published
2025-01-01
DOI
https://doi.org/10.2343/geochemj.gj25007
Primary Topic
X-ray Spectroscopy and Fluorescence Analysis
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Correction of X-ray tube aging effects and quantitative estimation of elemental composition using the ITRAX XRF core scanner: A case study of Japan Sea sediments

Masafumi Murayama, Ryuji Tada, Arisa Seki, Shunsuke Kurokawa
1 links
GEOCHEMICAL JOURNAL
X-ray Spectroscopy and Fluorescence Analysis
article

Correction of X-ray tube aging effects and quantitative estimation of elemental composition using the ITRAX XRF core scanner: A case study of Japan Sea sediments

Masafumi Murayama, Ryuji Tada, Arisa Seki, Shunsuke Kurokawa
article en

Abstract

The XRF core scanner, which facilitates non-destructive, high-speed, and high spatial resolution measurements, has recently been extensively used to measure element variability in sediments. Although identifying the precision of the measurement and concentration of the elements in the sample is essential to elucidating the variations recorded in the sediments, the element peak area count ratios are most commonly used to report the results of XRF core scanners in many previous studies, and element concentration is not discussed. This is mainly because quantitative calibration of element concentration necessitates numerous additional analyses of discrete samples. Because XRF core scanner measurements are often conducted before discrete sample analysis, if approximate element concentrations in the sediment can be estimated from XRF core scanner data just after measurement, sampling intervals should be determined for further analysis.In this study, we developed a new database on the aging of X-ray tubes, and propose a method for estimating element concentrations in the wet sediments from XRF core scanner measurement without any additional measurement. The ITRAX XRF core scanner at Kochi University, the first ITRAX installed in Japan in 2014, was used in this study. Mo X-ray tubes were used at 30 kV, 55 mA with XRF exposure time of 10 s or 32 s. First, the aging effect of the X-ray tube was monitored and its correction method was proposed. Subsequently, the precision of the ITRAX XRF measurements was evaluated and the relationship between the element peak area count by ITRAX and element concentration was analyzed. Finally, we propose a new method for applying these databases constructed in this study to any results measured by ITRAX XRF core scanner at Kochi University. Using this method, anyone can estimate approximate element composition of the sediment from their XRF core scanner data obtained at Kochi University without any additional measurement. This method even enables estimation of element concentrations of sediments just after the ITRAX measurement.

GEOCHEMICAL JOURNALVol. 59(4)
Shinshu University (JP), Fukada Geological Institute (JP), The University of Tokyo (JP), Kōchi University (JP), Chiba Institute of Technology (JP), Cosmo Oil (Japan) (JP)
Fujiwara Natural History Foundation, Japan Society for the Promotion of Science
Life below water
Openalex Percentile: Top 79%
X-ray Spectroscopy and Fluorescence Analysis
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.