Cardiac troponin detection in one drop of saliva or urine for 10-minute assessment of acute myocardial infarction

Acute myocardial infarction (AMI) can be fatal with sudden death, heart failure, and cardiac rupture. However, it is a challenge to achieve point-of-care testing (POCT) for cardiac troponin T (CTnT), an acute myocardial infarction biomarker, using conventional ELISA because it requires professional expertise and involves time-consuming steps. Herein, we propose a printed immunosensing photonic crystal biochip (PCB), which enables the sensitive, convenient, and non-invasive detection of CTnT. The fluorescence enhancement property of the PCB enables detection of CTnT within 10 minutes, with a limit of detection of 0.25 pg mL<sup>-1</sup>, compared to the conventional ELISA (2.5 h, 30 pg mL<sup>-1</sup>). Owing to the hydrodynamic enrichment of the PCB, 10 μL of sample can be measured, minimizing interference from impurities in body fluids and human sampling errors. We used 100 clinical samples, including serum, saliva, and urine, to verify the reliability of the PCB assay; compared with ELISA, the PCB assay exhibited consistencies of 0.998, 0.999, and 0.999, respectively. Notably, monitoring the saliva of a patient with an AMI history using the PCB assay enables rapid therapeutic intervention within 30 minutes. The PCB was first used for non-invasive body fluid testing to provide reliable results for clinical diagnosis, with the potential to provide AMI health promotion.

Authors

Institutions

Publication Details

Journal
Materials Horizons
Published
2025-01-01
DOI
https://doi.org/10.1039/d5mh00525f
Citations
3
Primary Topic
Advanced Biosensing Techniques and Applications
Type
article
Field-Weighted Citation Impact
2.42

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Cardiac troponin detection in one drop of saliva or urine for 10-minute assessment of acute myocardial infarction

Zewei Lian, Hua Shen, Meng Su, Zhiyu Tan et al.
3 citations
Materials Horizons
Advanced Biosensing Techniques and Applications
2.42
article

Cardiac troponin detection in one drop of saliva or urine for 10-minute assessment of acute myocardial infarction

Zewei Lian, Hua Shen, Meng Su, Zhiyu Tan, Jie Zhang, Yaqun Yu, Yundai Chen, Tong Ren, Nan Cheng, Xuelian Wu, Shengli Jiang
article en
3 citations

Abstract

Acute myocardial infarction (AMI) can be fatal with sudden death, heart failure, and cardiac rupture. However, it is a challenge to achieve point-of-care testing (POCT) for cardiac troponin T (CTnT), an acute myocardial infarction biomarker, using conventional ELISA because it requires professional expertise and involves time-consuming steps. Herein, we propose a printed immunosensing photonic crystal biochip (PCB), which enables the sensitive, convenient, and non-invasive detection of CTnT. The fluorescence enhancement property of the PCB enables detection of CTnT within 10 minutes, with a limit of detection of 0.25 pg mL<sup>-1</sup>, compared to the conventional ELISA (2.5 h, 30 pg mL<sup>-1</sup>). Owing to the hydrodynamic enrichment of the PCB, 10 μL of sample can be measured, minimizing interference from impurities in body fluids and human sampling errors. We used 100 clinical samples, including serum, saliva, and urine, to verify the reliability of the PCB assay; compared with ELISA, the PCB assay exhibited consistencies of 0.998, 0.999, and 0.999, respectively. Notably, monitoring the saliva of a patient with an AMI history using the PCB assay enables rapid therapeutic intervention within 30 minutes. The PCB was first used for non-invasive body fluid testing to provide reliable results for clinical diagnosis, with the potential to provide AMI health promotion.

Materials HorizonsVol. 12(19)
Chinese Academy of Sciences (CN), Chinese PLA General Hospital (CN), Beijing National Laboratory for Molecular Sciences (CN), Institute of Chemistry (CN)
National Natural Science Foundation of China, Ministry of Science and Technology of the People's Republic of China, Ministry of Science and Technology, National Key Research and Development Program of China, Beijing Science and Technology Planning Project
Good health and well-being
Openalex Percentile: Top 11%
Advanced Biosensing Techniques and Applications
2.42
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.