Development of a multiparameter phantom system for multispectral imaging

Multispectral imaging enables the visualization of physiological parameters, such as oxygen saturation (SO<sub>2</sub>) and lipid distribution. Assessing the performance of such imaging techniques necessitates the development of tissue-mimicking phantoms or phantom systems. However, the applicability of existing approaches is constrained by their reliance on homogeneous, unlayered tissue phantom and the absence of traceable reference standards. We propose a multiparameter phantom system featuring bidirectionally tunable SO<sub>2</sub> and a multilayered configuration comprising vascular, mucosal, and lipid layers. A blood gas analyzer was integrated as a calibration reference standard for SO<sub>2</sub> to ensure traceability. Two multispectral endoscopic imaging methods were validated using the phantom system. This cost-effective platform addresses the current gap in standardized evaluation of endoscopic devices for physiological parameter visualization.

Authors

Institutions

Publication Details

Journal
Biomedical Optics Express
Published
2025-06-25
DOI
https://doi.org/10.1364/boe.563392
Citations
1
Primary Topic
Optical Imaging and Spectroscopy Techniques
Type
article
Field-Weighted Citation Impact
1.36
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Development of a multiparameter phantom system for multispectral imaging

Wang We, Ji Qi, Qing Yang, Xiaopeng Chen et al.
1 citations
Biomedical Optics Express
Optical Imaging and Spectroscopy Techniques
1.36
article

Development of a multiparameter phantom system for multispectral imaging

Wang We, Ji Qi, Qing Yang, Xiaopeng Chen, Liqiang Wang, Xingjun Gao
article en
1 citations

Abstract

Multispectral imaging enables the visualization of physiological parameters, such as oxygen saturation (SO<sub>2</sub>) and lipid distribution. Assessing the performance of such imaging techniques necessitates the development of tissue-mimicking phantoms or phantom systems. However, the applicability of existing approaches is constrained by their reliance on homogeneous, unlayered tissue phantom and the absence of traceable reference standards. We propose a multiparameter phantom system featuring bidirectionally tunable SO<sub>2</sub> and a multilayered configuration comprising vascular, mucosal, and lipid layers. A blood gas analyzer was integrated as a calibration reference standard for SO<sub>2</sub> to ensure traceability. Two multispectral endoscopic imaging methods were validated using the phantom system. This cost-effective platform addresses the current gap in standardized evaluation of endoscopic devices for physiological parameter visualization.

Biomedical Optics ExpressVol. 16(9)
MACOM (United States) (US), University of Electronic Science and Technology of China (CN), Zhejiang University (CN), University of Hong Kong (HK)
Openalex Percentile: Top 19%
Optical Imaging and Spectroscopy Techniques
1.36
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.