Correction of thermal inertia on scanning calorimetry

• Correction methods of the thermal inertia on the heat flow curve are reviewed. • The sample influence for thermal inertia on FDSC is discussed. • The use of the sample for measurement of the characteristic time is proposed. • The determination of thermal inertia by temperate modulated DSC is introduced. In differential scanning calorimetry (DSC), the thermal inertia (TI) of the calorimeter and the sample influences the shape of the measurement curve. This affects the evaluation of characteristic temperatures and kinetic quantities. Therefore, the significant influence of TI must be corrected. In conventional DSC, the TI is essentially given by the device and the crucible and the sample contribution can often be neglected. However, the situation is different with chip-based fast scanning differential calorimetry (FDSC), where the DSC furnace is the active zone of the sensor membrane. Due to the larger dynamic range in FDSC and the possible similarity between the heat capacity of the active zone of the sensor and the sample, the sample must be considered when determining and correcting TI. Various correction methods that can be used for DSC and FDSC are discussed here

Authors

Institutions

Publication Details

Journal
Thermochimica Acta
Published
2025-06-24
DOI
https://doi.org/10.1016/j.tca.2025.180062
Citations
4
Primary Topic
Thermal and Kinetic Analysis
Type
article
Field-Weighted Citation Impact
2.07
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Correction of thermal inertia on scanning calorimetry

Jürgen E. K. Schawe
4 citations
Thermochimica Acta
Thermal and Kinetic Analysis
2.07
article

Correction of thermal inertia on scanning calorimetry

Jürgen E. K. Schawe
article en
4 citations

Abstract

• Correction methods of the thermal inertia on the heat flow curve are reviewed. • The sample influence for thermal inertia on FDSC is discussed. • The use of the sample for measurement of the characteristic time is proposed. • The determination of thermal inertia by temperate modulated DSC is introduced. In differential scanning calorimetry (DSC), the thermal inertia (TI) of the calorimeter and the sample influences the shape of the measurement curve. This affects the evaluation of characteristic temperatures and kinetic quantities. Therefore, the significant influence of TI must be corrected. In conventional DSC, the TI is essentially given by the device and the crucible and the sample contribution can often be neglected. However, the situation is different with chip-based fast scanning differential calorimetry (FDSC), where the DSC furnace is the active zone of the sensor membrane. Due to the larger dynamic range in FDSC and the possible similarity between the heat capacity of the active zone of the sensor and the sample, the sample must be considered when determining and correcting TI. Various correction methods that can be used for DSC and FDSC are discussed here

Thermochimica ActaVol. 751
ETH Zurich (CH)
Affordable and clean energy
Openalex Percentile: Top 13%
Thermal and Kinetic Analysis
2.07
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.