Standardisation of iodine-123 and participation in the comparison BIPM.RI(II)-K4.I-123

Iodine-123 was standardised by the 4π(e<sub>A</sub>,X)-γ live-timed anti-coincidence counting technique, with liquid scintillation (LS) counting applied for the 4π channel. Seven LS sources were counted over 4 days. To trace the decay of impurities <sup>125</sup>I, <sup>121</sup>Te and <sup>123m</sup>Te, one source was counted for an additional 9 times over 100 days. Fitting the extended dataset to the decay function for the component radionuclides provided impurity correction factors. The recorded data were corrected for background, live time, impurities and decay, and plotted as N<sub>LS</sub>N<sub>γ</sub>N<sub>C</sub> vs (N<sub>γ</sub>N<sub>C</sub>-1). Weighted least squares regression and extrapolation to (N<sub>γ</sub>N<sub>C</sub>-1)=0 provided the source activity. Data were fitted with a 3rd order polynomial with the 2nd order coefficient = zero to account for curvature present in the data. A 1st order polynomial fit provided an activity concentration that was 1.02 % higher. This primary standard was developed as part of ANSTO's participation in the BIPM. RI(II)-K4. I-123 key comparison, using the Transfer Instrument of the International Reference System (SIRTI). The result that was submitted for participation in the inter-comparison was 0.85 % higher than the result reported in this manuscript. With the latest re-evaluation, the standard uncertainty increased from 0.36 % to 0.53 %, mostly due to the non-linear extrapolation of the detection efficiency curve.

Authors

Institutions

Publication Details

Journal
Applied Radiation and Isotopes
Published
2025-06-24
DOI
https://doi.org/10.1016/j.apradiso.2025.112012
Primary Topic
Radioactive Decay and Measurement Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Standardisation of iodine-123 and participation in the comparison BIPM.RI(II)-K4.I-123

Melek Zarifi, Christine M.B. Keevers, Emily L. Clark, Angus H.H. Bowan et al.
Applied Radiation and Isotopes
Radioactive Decay and Measurement Techniques
article

Standardisation of iodine-123 and participation in the comparison BIPM.RI(II)-K4.I-123

Melek Zarifi, Christine M.B. Keevers, Emily L. Clark, Angus H.H. Bowan, W.M. van Wyngaardt, C. Michotte, M. L. Smith
article en

Abstract

Iodine-123 was standardised by the 4π(e<sub>A</sub>,X)-γ live-timed anti-coincidence counting technique, with liquid scintillation (LS) counting applied for the 4π channel. Seven LS sources were counted over 4 days. To trace the decay of impurities <sup>125</sup>I, <sup>121</sup>Te and <sup>123m</sup>Te, one source was counted for an additional 9 times over 100 days. Fitting the extended dataset to the decay function for the component radionuclides provided impurity correction factors. The recorded data were corrected for background, live time, impurities and decay, and plotted as N<sub>LS</sub>N<sub>γ</sub>N<sub>C</sub> vs (N<sub>γ</sub>N<sub>C</sub>-1). Weighted least squares regression and extrapolation to (N<sub>γ</sub>N<sub>C</sub>-1)=0 provided the source activity. Data were fitted with a 3rd order polynomial with the 2nd order coefficient = zero to account for curvature present in the data. A 1st order polynomial fit provided an activity concentration that was 1.02 % higher. This primary standard was developed as part of ANSTO's participation in the BIPM. RI(II)-K4. I-123 key comparison, using the Transfer Instrument of the International Reference System (SIRTI). The result that was submitted for participation in the inter-comparison was 0.85 % higher than the result reported in this manuscript. With the latest re-evaluation, the standard uncertainty increased from 0.36 % to 0.53 %, mostly due to the non-linear extrapolation of the detection efficiency curve.

Applied Radiation and IsotopesVol. 225
Australian Nuclear Science and Technology Organisation (AU), Bureau international des poids et mesures (FR)
Openalex Percentile: Top 79%
Radioactive Decay and Measurement Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Standardisation of iodine-123 and participation in the comparison BIPM.RI(II)-K4.I-123 — Melek Zarifi, Christine M.B. Keevers, et al. · Applied Radiation and Isotopes (2025) | TGRS Research Map | TGRS