Consistency Analysis of Centiloid Values Across Three Commercial Software Platforms for Amyloid PET Quantification

<b>Objectives</b>: This study aimed to evaluate the consistency of Centiloid (CL) values calculated using three commercially available software platforms: BTXBrain (v1.1.2), MIM (v7.3.7), and SCALE PET (v2.0.1). <b>Methods</b>: A total of 239 patients who underwent amyloid PET/CT with either F-18 flutemetamol (FMM) or F-18 florbetaben (FBB) were retrospectively analyzed. CL values were calculated using BTXBrain, MIM, and SCALE PET. Linear regression, Passing-Bablok regression, and Bland-Altman analysis were performed to assess the agreement between CL values. Subgroup analyses were conducted for each radiotracer. CL values were compared according to visual interpretation status. <b>Results</b>: Strong correlations were observed between CL values derived from the three software platforms (R<sup>2</sup> > 0.95). However, Passing-Bablok regression revealed significant proportional bias, with CL values from BTXBrain being lower than others, and CL values from SCALE PET being higher than others as CL values increased. Bland-Altman plots visualized the proportional bias, particularly between BTXBrain and SCALE PET. Subgroup analyses by radiotracer showed similar results. CL values in visually positive scans were significantly higher than those in visually negative scans across all platforms. <b>Conclusions</b>: The three commercial software programs demonstrated high consistency in CL quantification. However, a notable systematic bias was observed. Further evaluation of various scanner effects and CL calculation methods is warranted to improve the consistency and reproducibility of CL quantification in clinical practice.

Authors

Institutions

Publication Details

Journal
Diagnostics
Published
2025-06-24
DOI
https://doi.org/10.3390/diagnostics15131599
Citations
3
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
4.08

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Consistency Analysis of Centiloid Values Across Three Commercial Software Platforms for Amyloid PET Quantification

Narae Lee, Hyukjin Yoon, Yoo Hyun Um, Woo Hee Choi
3 citations
Diagnostics
Medical Imaging Techniques and Applications
4.08
article

Consistency Analysis of Centiloid Values Across Three Commercial Software Platforms for Amyloid PET Quantification

Narae Lee, Hyukjin Yoon, Yoo Hyun Um, Woo Hee Choi
article en
3 citations

Abstract

<b>Objectives</b>: This study aimed to evaluate the consistency of Centiloid (CL) values calculated using three commercially available software platforms: BTXBrain (v1.1.2), MIM (v7.3.7), and SCALE PET (v2.0.1). <b>Methods</b>: A total of 239 patients who underwent amyloid PET/CT with either F-18 flutemetamol (FMM) or F-18 florbetaben (FBB) were retrospectively analyzed. CL values were calculated using BTXBrain, MIM, and SCALE PET. Linear regression, Passing-Bablok regression, and Bland-Altman analysis were performed to assess the agreement between CL values. Subgroup analyses were conducted for each radiotracer. CL values were compared according to visual interpretation status. <b>Results</b>: Strong correlations were observed between CL values derived from the three software platforms (R<sup>2</sup> > 0.95). However, Passing-Bablok regression revealed significant proportional bias, with CL values from BTXBrain being lower than others, and CL values from SCALE PET being higher than others as CL values increased. Bland-Altman plots visualized the proportional bias, particularly between BTXBrain and SCALE PET. Subgroup analyses by radiotracer showed similar results. CL values in visually positive scans were significantly higher than those in visually negative scans across all platforms. <b>Conclusions</b>: The three commercial software programs demonstrated high consistency in CL quantification. However, a notable systematic bias was observed. Further evaluation of various scanner effects and CL calculation methods is warranted to improve the consistency and reproducibility of CL quantification in clinical practice.

DiagnosticsVol. 15(13)
The Catholic University of Korea St. Vincent's Hospital (KR), Catholic University of Korea (KR)
National Research Foundation of Korea
Openalex Percentile: Top 6%
Medical Imaging Techniques and Applications
4.08
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.